Receiver Mode
4.44
Operating Manual 1302.6163.12 - 03
RUN
FINAL MEAS
Sequence for
RUN FINAL MEAS
The
RUN FINAL MEAS
softkey starts the final measurement sequence as described
below. The
HOLD FINAL MEAS
submenu is called.
Dependent on the selected measurement mode (
AUTOMATIC FINAL
or
INTER
ACTIVE
) for the final measurement, two different submenus of the
HOLD FINAL
MEAS
menu are displayed.
The detectors used for the final measurement replace those used in the list or for the
prescan.
The out-of-limit condition is shown by a plus sign preceding the values in the
DELTA
LIMIT
column
(see table below).
Remote command:
--
HOLD
FINAL MEAS
Dependent on the selected measurement mode (
AUTOMATIC FINAL
or
INTER
ACTIVE
) for the measurement, two different submenus of the
HOLD FINAL MEAS
menu are displayed.
The
HOLD FINAL MEAS
softkey interrupts the automatic run of the final measure-
ment.
The
CONTINUE FINAL MEAS
submenu appears.
Sequence for
AUTOMATIC FINAL
With the final measurement halted all receiver settings can be modified for example
for examining the signal in detail.
The mode of the measurement (automatic or interactive) can again be selected.
The final measurement can be started with the
MEASURE
softkey.
Remote command:
--
AUTOMATIC
FINAL
The
AUTOMATIC FINAL
softkey selects the automatic mode for the final measure-
ment (see above).
Remote command:
:FME:AUTO ON
The final measurement function is only available in the manual operation. For
operation under program control it is preferable to read the prescan results and
the data-reduced peak list, if any, from the R&S ESU with the controller and carry
out the individual measurements with the controller. This considerably makes the
control of the interactive mode easier.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...