Operating Manual 1302.6163.12 - 03
6.87
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.5.5
CALCulate:MATH Subsystem
The CALCulate:MATH subsystem allows to process data from the SENSe-subsystem in numeric expres-
sions.
The measurement windows are selected by CALCulate1 (screen A) and CALCulate2 (screen B).
CALCulate<1|2>:MATH[:EXPression][:DEFine]
(<expr>)
This command defines the mathematical expression for relating traces to trace1.
The
command switches the mathematical relation of traces on or off.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Parameter:
<expr>::= 'OP1 - OP2'
OP1 ::= TRACE1
OP2 ::= TRACE2 | TRACE3
Example:
"CALC1:MATH (TRACE1 - TRACE2)"
'Selects the subtraction of trace 1 from trace 2 in screen A.
"CALC2:MATH (TRACE1 - TRACE3)"
'Selects the subtraction of trace 1 from trace 3 in screen B.
Characteristics:
*RST value: -
SCPI: conform
Mode:
A-T
CALCulate<1|2>:MATH:POSition
-100PCT to 200PCT
This command defines the position of the result of the trace mathematics in the selected measurement
window. The indication is in % of the screen height, with 100% corresponding to the upper diagram
border.
Example:
"CALC:MATH:POS 50PCT"
'Sets the position in screen A to the horizontal diagram center.
Characteristics:
*RST value: 50 %
SCPI: device-specific
Mode:
A-T
CALCulate<1|2>:MATH:STATe
ON | OFF
This command switches the mathematical relation of traces on or off.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Example:
"CALC:MATH:STAT ON"
'Switches on the trace mathematics in screen A.
Characteristics:
*RST value: OFF
SCPI: conform
Mode:
A-T
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...