Operating Manual 1302.6163.12 - 03
4.203
R&S ESU
Instrument Functions
Analyzer Mode
PERCENT
MARKER
If the CCDF function is active, the
PERCENT MARKER
softkey allows to position
marker 1 by entering a probability value. Thus, the power which is exceeded with a
given probability can be determined very easily.
If marker 1 is in the switched-off state, it will be switched on automatically.
Remote command:
CALC:MARK:Y:PERC 0...100%
RES BW
The
RES BW
softkey sets the resolution bandwidth in the menu
STATISTIC FUNC-
TION
directly without switching to the corresponding menu (
BW
). The function of the
softkey is identical to the softkey
RES BW MANUAL
in the
BW
menu.
For correct measurement of the signal statistics the resolution bandwidth has to be
wider than the signal bandwidth in order to transmit the actual peaks of the signal
amplitude correctly. Video bandwidth is set to 10 MHz automatically with a statistic
function switched on.
Remote command:
BAND 3 MHz
NO OF
SAMPLES
The
NO OF SAMPLES
softkey sets the number of power measurements taken into
account for the statistics.
Please note that the overall measurement time is influenced by the number of sam-
ples selected as well as by the resolution bandwidth set up for the measurement as
the resolution bandwidth directly influences the sampling rate.
Remote command:
CALC:STAT:NSAM <value>
SCALING
The
SCALING
softkey opens a submenu that allows changing the scaling parame-
ters for both the x- and the y-axis.
X-AXIS REF
LEVEL
The
X-AXIS REF LEVEL
softkey changes the level settings of the instrument and
sets the maximum power to be measured.
The function is identical to softkey
REF LEVEL
in
AMPT
menu.
For the
APD
function this value is mapped to the right diagram border. For the
CCDF
function there is no direct representation of this value on the diagram as the
x-axis is scaled relatively to the
MEAN POWER
measured.
Remote command:
CALC:STAT:SCAL:X:RLEV <value>
X-AXIS RANGE
The
X-AXIS RANGE
softkey changes the level range to be covered by the statistics
measurement selected.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...