R&S ESU
Instrument Functions
Analyzer Mode
4.204
Operating Manual 1302.6163.12 - 03
The function is identical to softkey
RANGE LOG MANUAL
in
AMPT
menu.
Remote command:
CALC:STAT:SCAL:X:RANG <value>
Y-UNIT %/ABS
The softkey
Y-UNIT %/ABS
defines the scaling type on the y-axis. The default case
are the absolute probability. This can be changed to percent values. The softkeys
Y-
AXIS MIN
and
Y-AXIS MAX
are using values based on the
Y-UNIT
setting.
Remote command:
CALC:STAT:SCAL:Y:UNIT PCT | ABS
The 0.01%, 0.1%, 1% and 10% value from the CCDF measurement are shown in
the bottom screen half. Those values can also queried via remote:
Remote command:
CALC:STAT:CCDF:X? P0_01 | P0_1 |
P1 | P10
Y-AXIS MAX
VALUE
The
Y-AXIS MAX VALUE
softkey defines the upper limit of the displayed probability
range.
Values on the y-axis are normalized which means that the maximum value is 1.0. As
the y-axis scaling has a logarithmic axis the distance between max and min value
must be at least one decade.
Remote command:
CALC:STAT:SCAL:Y:UPP <value>
Y-AXIS MIN
VALUE
The
Y-AXIS MIN VALUE
softkey defines the lower limit of the displayed probability
range.
As the y-axis scaling has a logarithmic axis the distance between max and min value
must be at least one decade. Valid values are in the range 0 < value < 1.
Remote command:
CALC:STAT:SCAL:Y:LOW <value>
ADJUST
SETTINGS
The
ADJUST SETTINGS
softkey optimizes the level settings of the R&S ESU
according to the measured peak power in order to gain maximum sensitivity of the
instrument.
The level range is adjusted according to the measured difference between peak and
minimum power for APD measurement and peak and mean power for CCDF mea-
surement in order to obtain maximum power resolution.
Additionally the probability scale is adapted to the selected number of samples.
Remote command:
CALC:STAT:SCAL:AUTO ONCE
DEFAULT
SETTINGS
The
DEFAULT SETTINGS
softkey resets the x- and y-axis scalings to their PRESET
values.
• x-axis ref level: -20 dBm
• x-axis range APD: 100 dB
• x-axis range CCDF: 20 dB
• y-axis upper limit: 1.0
• y-axis lower limit: 1E-6
Remote command:
CALC:STAT:PRES
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...