Section IV Care
Key points for trouble-free operation:
•
Rinse vacuum lines daily with the recommend
-
ed quantity of liquid and line cleaner.
•
Check air filter on schedule.
•
Exchange the pump head on schedule.
Foam may cause damage to the pump. Avoid
interruptions by insuring your cleaner is truly
“non-foaming”. Test by shaking the mixed
solution. True “non-foamers” will be bubble free.
CAUTION
Prior to performing any maintenance, proper
precautions should be taken to reduce the pos
-
sibility of contact with infectious substances.
WARNING
Maintenance Overview
Preventative Maintenance Schedule
IMPORTANT: Use of line cleaners outside of
this pH range will void the warranty.
CAUTION
Use of SlugBuster
TM
vacuum line cleaner is
highly recommended. SlugBuster is available
in powder and liquid forms, which can be or
-
dered through any authorized RAMVAC dealer.
NOTICE
SlugBuster Ordering Information
RAMVAC
Part No.
Contents
900143
Powder Starter Pack: Mixing jug, two 40 oz.
containers of SlugBuster.
900146
Powder 6 Pack: six 40 oz. containers of
SlugBuster (makes 480 gallons of solution)
900241
Liquid Single Gallon: 1 gallon of SlugBuster
concentrate and hand pump dispenser
900243
Liquid Starter Pack: Mixing jug, 1 gallon of Slug
-
Buster concentrate and hand pump dispenser
900246
Liquid 4 Pack Case: 4 single gallons of SlugBuster
concentrate and 4 hand pump dispensers
Cleaning the Vacuum System
NOTE
:
The Vacuum System should be cleaned daily.
NOTE
:
Check the treatment room solids separator
routinely and clean when dirty.
1.
Just before turning off the RAMVAC, rinse vac
-
uum lines first with hot water - approximately
one quart through each high volume line and a
few ounces through each saliva ejector line.
2.
Aspirate a few ounces of a dental vacuum line
cleaner through each vacuum line.
SlugBuster
TM
is highly recommended.
After Surgical Procedures
1.
Aspirate a few ounces of an appropriate vacu
-
um line cleaner, such as SlugBuster
TM
, through
the lines.
For Overhead Plumbing
1.
Be sure to allow air to follow liquids before
closing vacuum valves.
Cleaners
Cleaners should have these qualities:
•
Non-foaming
•
Deodorizing
•
Sanitizing
•
pH between 6 and 10
Daily - Rinse vacuum lines
See "Cleaning the Vacuum System"
Every 1,000 hours* - Inspect air filters
See "Changing the Air Filters"
Every 12,000 Hours* - Exchange pump head
See "Pump Head Exchange"
*
The "Maintenance" LED on the E2 Control will
illuminate every 1,000 hours.
*
To reset the "Maintenance" LED after
maintenance has been performed, press and
hold the "Reset" button for 10 seconds until the
"Maintenance" LED begins blinking.