NSTALLATION
14
R22
1/A
- R221
/A DHHS -
Inst
all
ation
Ralco
srl
E
LECTRICAL
C
ONNECTION
Supply and signals to the collimator must be to 2007/47/CE standards. Devices that supply to
the collimator must therefore feature double or reinforced insulation as provided by the General
Standard on Electro-medical Equipment CEI62-5 +A2 (EN 60601-1).
The device is electrostatic sensitive, consequently all the relating safety standards must
be complied with
.
Power Supply Connection
• Remove the part of cover to access the terminal, see Chapter-
C
OVER
R
EMOVAL
.
• Connect supply cables to the relative collimator terminals and earth on the screw marked
with the relating symbol:
. Use the protective earth cable with green/yellow
insulation.
• In collimators with a free cable stop, remove the cable stop, connect the supply cables,
remount and then secure the cable stop.
T
HE
WIRING
DIAGRAM
INCLUDED
IN
THIS
CHAPTER
REFERS
TO
THE
STANDARD
PRODUCT
.
I
T
IS
THE
RESPONSIBILITY
OF
THE
CUSTOMER
WHO
HAS
REQUESTED
AN
ELECTRIC
CUSTOMIZATION
,
TO
ENSURE
THAT
AN
ELECTRIC
DIAGRAM
RELATING
TO
THE
CUSTOMIZATION
HAS
BEEN
PROVIDED
WITH
THE
DOCUMENTATION
.
T
HE
DEVICE
MUST
BE
EXCLUSIVELY
CONNECTED
TO
POWER
NETWORK
,
WITH
EARTH
PROTEC
-
TION
,
IN
ORDER
TO
AVOID
A
RISK
OF
ELECTRICAL
SHOCK
.
T
HE
DEVICE
IS
PROJECTED
TO
OPERATE
WITH
A
PERMANENT
POWER
SUPPLY
PRESENT
SO
THE
PROCEDURE
OF
SWITCHING
OFF
THE
COLLIMATOR
IS
NOT
FORESEEN
.
N
O
RISK
OR
DEVICE
DAMAGE
WILL
OCCUR
IF
THE
MACHINE
IS
ACCIDENTLY
SWITCHED
OFF
.
C
OLLIMATOR
MUST
BE
SUPPLIED
AS
SPECIFIED
,
SEE
C
HAPTER
-
S
PECIFICATIONS
.
THE
SUPPLY
MUST
COME
FROM
A
SEPARATE
SOURCE
FROM
THE
POWER
NETWORK
THROUGH
DOUBLE
INSULATION
OR
REINFORCED
INSULATION
AND
WITH
LIMITED
CURRENT
.
TRANSFORMER
CHAR
-
ACTERISTICS
MUST
CONFORM
TO
THE
REQUIREMENTS
OF
STANDARD
IEC 60601-1.
CABLES
AND
TERMINALS
USED
FOR
THE
INTERNAL
CONNECTION
OF
THE
COLLIMATOR
MUST
BE
SUITABLE
FOR
OPERATION
AT
TEMPERATURES
OF
70°
C
AND
COLLIMATOR
CURRENT
ABSORPTION
.
Summary of Contents for MTR 221/A