MTR221/A - MTR221/A DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
Modello/Model R221/A - R221/A DHHS
D
ESCRIPTION
6
R22
1/A
- R221
/A DHHS -
D
escription
Ralco
srl
DESCRIPTION
Multilayer, square field, manual collimation system with a lightweight and compact
design intended for installation on mobile and stationary X-ray equipment. This device
has been designed and manufactured for skeletal investigations and ER applications.
The X-ray field is defined by six pairs of shutters, four of which are lead-lined. The six pairs of
shutters move perpendicularly within the X-ray field. Two pairs of brass shutters are located
near the focus, two are located near the entrance window and two are located near the exit
window of the X-ray beam from the collimator. The latter shutters serve to accurately define the
X-ray field edges.
Shutter movements are manual, controlled by two knobs on the collimator front panel.
C
HARACTERISTICS
Mounting Plane
at 80 mm (3.14") from the focus.
Continuous Film Coverage
from Min: 00 x 00 cm to Max: 48 x 48 cm at 100 cm (40") SID.
Maximum Radiation Leakage:
150 kVp - 4 mA.
Accessory Guides
are used for accessories and additonal filtration.
Retractable Tape
mounted on a radiological unit, measures the distance between the
focus and the patient.
Halogen Lamp
providing high luminosity for the X-ray field projection.
Minimum Inherent Filtration
2 mm aluminium equivalent.
GC-Led-5A
timer board for light source supply and operation.
Fig. R221/A- R221/A DHHS Front Panel with Exterior Dimensions
This collimator may have the following optional items; a detailed description is provided in the
chapter
O
PTIONAL
I
TEMS
.
Summary of Contents for MTR 221/A
Page 66: ...info_cl fm...