RainDrop Sense Operator’s Manual
LCN 50-04343 Rev. D
7-5
6.
Place the tip of the syringe over the fill port and begin to turn the entire syringe
clockwise. The syringe screws onto the fill port to create a tight fit.
Figure 7-3: Syringe Connected to the Fill Port
7.
Continue to turn the syringe until it fits snugly.
Caution:
Do not overtighten the syringe onto the fill port.
8.
Press the end of the syringe and inject its contents until the syringe is empty.
9.
Unscrew the empty syringe and discard it in a lab-approved solid waste stream. When
you are done refilling the reservoir, the oil low indicator light turns off.
10.
Replace the cover on the fill port. Do not overtighten.