Instructions For Use
3
3.
Remove the plug from the reservoir
priming port and insert injection tee
tip into port. Turn the ball valve to
the open position.
4.
press the start/
pause button and
allow product to
flow until there
are no visible air
bubbles within the
tubes.
5.
After completing the priming process, press
pause, then switch setting back to injection
mode by pressing the left or right arrows until
screen display changes over to injection mode.
IV.
Calibrate
1.
Once in purge mode, press the up button to switch the
device to calibration mode.
2.
Once in calibration mode, insert injection tip into a gradu-
ated cylinder and press start to inject the ‘10 ml’ calibration
dose (a 10 ml graduated cylinder has been provided with the
device for this purpose). The device will distribute a dose
that may be above or below the intended 10 ml dose.
Press the left or right arrows until screen display
changes over to purge mode.
10mL
Perform calibration at start of day and when changing products.