background image

Technical Note TN-204    

04/17/vk 

 

Honeywell Internal

 

RAE Systems by Honeywell    877-723-2878     raesystems.com 

 

Tube capacitance can be extended significantly while using the 

instrument’s low pump speed (LPS) instead of its high pump 

speed as default setting. Test conditions are the same as above. 

 

Note: Previous results reinforce our strong recommendation to 

only open RAE-Sep Tubes right before performing a benzene-

specific  measurement.  Other approaches  could lead to false 

positive results. 

Temperature Effects 

Temperature variation affects RAE-Sep Tube performance. De-

creasing temperature leads to the higher tube absorption, but slows 

down chemical transformation. As a result, the tube may not show a 

color change, but the tube fully or partially loses its capacitance. To 

avoid confusion, follow the recommendation of “One tube, one 

measurement.” Note: Calibration should be performed at the same 

temperature as the measurement because time necessary for sam-

pling changes with temperature. 

 

Tube Capacitance 

Tube capacitance in this case is defined as: MultiRAE Benzene 

with a RAE-Sep Tube exposed to the gases in the table for a 

limited time until the reading reached 0.5 ppm. 

Concentration vs. Exposure Time 

 

Gas 

Gas concentration, 

ppm 

Capacitance, 

minutes 

Toluene 

100 

≥10 

i-Butylene 

100 

≥10 

i-Butylene 

1000 

Methane 

25000 

No response  

n-Hexane 

2400 

10 sec 

n-Octane 

500 

1* 

n-Pentane 

3500 

7 sec 

n-Propane 

11000 

No response  

* Reading 1 ppm after 1 minute exposure 

Potential Cross-Sensitivity 

Gas Cross-

sensitivity  

Gas concentration, 

ppm 

Benzene 

Response, ppm 

Pentane 

1700 

0.1 

Propane 

5500 

i-Butane 

4500 

0.5 

i-Butylene 

100 

0.6 

i-Butylene 

1000 

0.8 

Methane 

12500 

Octane 

250 

0.3 

Hexane 

1200 

0.6 

Hexane 

100 

Butadiene 1,3 

2.5 

0.5 

Toluene 

50 

Acetone 

200 

0.1 

 

Benzene Absorption by RAE-Sep Tubes 

There is a small effect of the benzene absorbtion by the RAE-

Sep  Tube that potentially could slightly decrease a benzene 

reading, but it doesn’t exceed 50 to 60 ppb, which is below the 

instrument’s resolution of 0.1 ppm (100ppb).  

 

 

 

 

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

0

2

4

6

8

Re

ad

in

g,

 p

pm

Numbers of measurements

Octane, response vs. numbers 

of measurements

HPS-1
LPS-1
HPS-2
LPS-2

0

50

100

150

200

250

300

110

149

225

328

Rea

ding, ppb

Benzene concentration, ppb

Effect of tube on Benzene absorption

Without tube

With tube

Reviews: