Rae TN-204 User Manual Download Page 2

Technical Note TN-204    

04/17/vk 

 

Honeywell Internal

 

RAE Systems by Honeywell    877-723-2878     raesystems.com 

 

Some measurable parameters of the instrument are presented be-

low. Refer to the MultiRAE Benzene datasheet for comprehensive 

sensor specification data.

 

Sensor/Mode 

Range 

Reso-

lution 

9.8eV for use in TVOC mode  10 ppb to 2,000 ppm 

10 ppb 

9.8eV for use in Benzene 
mode (with RAE-Sep Tube 
Cartridge) 

0.1 to 200 ppm 

0.1 ppm 

 

MULTIRAE BENZENE-SPECIFIC MEASUREMENT 

PRINCIPLE

  

In Benzene mode, the monitor is designed to make specific ben-

zene measurements. This is achieved via these two approaches: 

1.

 

The monitor uses a lower-energy, 9.8eV PID lamp. This al-

lows excluding from cross-contamination all the VOCs with 

ionization energy (IE) above 9.8eV. 

2.

 

RAE-Sep tubes eliminate most VOCs. Two major effects 

are present in them: 

 

Gas absorption in the tube. This is usually a fast pro-

cess, and it is reversible.  

 

Gas reaction (chemical) in the tube. This is a slower 

process, and it is not reversible. 

 

Note: Some VOCs can be eluted (removed) from the RAE-Sep 

Tube and cause false positive benzene readings. 

In addition, if the instruments are insensitive to some gases (for 

example, their IE is above 9.8eV or if they require high correction 

factors), the products of their chemical reactions in RAE-Sep 

Tubes may become sensitive and may distort the benzene read-

ing. A separate case is when gases don’t absorb in the RAE-

Sep Tube and are insensitive to the monitor equipped with a 

9.8eV lamp (for example, lower hydrocarbons such as methane, 

ethane, etc.). 

 

Example of Elution 

When the instrument in Benzene mode was exposed to 400 ppm 

of acetone for 60 seconds, the reading didn’t exceed 0.2 ppm. 

After 5 minutes  running in air, the  reading increased  and 

reached almost 30 ppm. 

 

 

Elution in Benzene mode with exposure to 400 ppm of  

acetone. 

Acetone didn’t convert in the RAE-Sep Tube within a short pe-

riod of time (slow chemical reaction) and further eluted with the 

air stream. 

Important: For the correct measurement and to avoid a false positive 

response, RAE-Sep Tubes should be used just one time. 

Humidity Effect 

Humidity has no effect on the response to benzene. However, 

high humidity affects the capacitance of the tubes to remove in-

terfering hydrocarbons, especially when they are in high con-

centrations.  

 

The figure below shows the response of the instrument to the 

mix of benzene (2 ppm) and octane (180 ppm). HPS-2 is the 

case when the tube was opened and the instrument run with a 

high pump speed (HPS) for about 5 minutes in ambient air with 

relative humidity (RH) of 75 to 80% before a benzene measure-

ment in Benzene mode was done. HPS-1 is the case when ben-

zene measurements were done right after opening the tube. 

 

Response of Benzene and Octane with different times of 

measurement after opening the tubes, HPS-1 and HPS-2. 

Each line represents a single tube, and each triangle repre-

sents a single measurement with that tube. 

 

0

5000

10000

15000

20000

25000

30000

35000

0

10

20

30

Re

sp

on

se

, p

pb

Time, min

Acetone eluate from Sep-Tube

ppb

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

0

2

4

6

8

Re

ad

in

g,

 p

pm

Numbers of measurements

Octane, response vs. numbers 

of measurements

HPS-1
HPS-2

Reviews: