MiniRAE 3000/ppbRAE 3000 Pocket Reference
31
When Span calibration is complete, you see this
message:
Span 1 is done!
Reading = 100.0 ppm
The instrument then exits Span calibration and shows
the Zero calibration menu on its display.
Note:
The reading should be very close to the span gas
value.