GammaRAE II

Addendum For

Firmware Version 2.00

Intrinsically Safe

Personal Radiation Monitor

Part Number 047-4105-100
Revision A, November 2005

Saudi Arabia :  [email protected]
U.A.E : [email protected]

Summary of Contents for GammaRAE II

Page 1: ...aRAE II Addendum For Firmware Version 2 00 Intrinsically Safe Personal Radiation Monitor Part Number 047 4105 100 Revision A November 2005 Saudi Arabia info_ksa magusintl com U A E info_uae magusintl...

Page 2: ...GammaRAE II Introduction This addendum covers changes in the GammaRAE II firmware from version 1 00 to version 2 00 and replaces the Operation section of the included manual...

Page 3: ...PEAK 6 MIN 6 DOSE 7 TIME 8 STAT 8 COMM 9 TEST 9 Programming Mode 10 Alarm Set 12 Datalog Set 14 Monitor Set 15 Back Lite 16 Buzz Lite 16 Vibrate mode 16 Temp Unit 17 Gamma Unit 17 Change Date 17 Chang...

Page 4: ...ions NORM Normal Operating Mode CAL Background calibration only in Search Mode PEAK Maximum radiation level detected since last cleared MIN Minimum radiation level detected since last cleared DOSE Acc...

Page 5: ...g Mode Switch back and forth between units by pressing the SET button Press the MODE button to step to the next function CAL Background calibration This option is only active if the GammaRAE II is in...

Page 6: ...e reading press MODE to return to the PEAK reading The PEAK value is also cleared when the detector is turned off Press the MODE button to step to the next function MIN The minimum radiation level det...

Page 7: ...step to the next function DOSE The accumulated radiation dosage since last cleared Press the SET button twice to clear the DOSE value If you press SET once but do not want to clear the DOSE reading pr...

Page 8: ...format can be set in Programming Mode On Run time from when the detector was turned on shown in hours and minutes 0000 00 Press the MODE button to step to the next function STAT Detector status BattV...

Page 9: ...lf test This consists of tests to make sure all alert functions are working properly This function is only enabled if the Diagnostic Flag was set at start up To set the Diagnostic Flag press and hold...

Page 10: ...the SET button to start the next test The LEDs are tested Press the SET button to start the next test The back light is tested Press the SET button to start the next test The BlueTooth radio is tested...

Page 11: ...atically move on to the next test Programming Mode Programming Mode is used to change alarm settings and detector settings units of measure time etc Note Changes to some parameters cause the detector...

Page 12: ...er Alarm Set Select Alarm Set to set the alarm mode or type and the detection limits for the alarm Parameter Used In Alarm Type Select Search or Safety Mode Gamma CF Search Mode Gamma Hi Safety Mode G...

Page 13: ...to step through the digits from left to right Use the SET button to increment each digit from 0 through 9 Select Save to save changes or Quit to discard changes To exitAlarm Set step through the opti...

Page 14: ...he time interval between datalog points You may set any time interval between 1 and 3600 seconds Press the MODE button to step through the digits from left to right Use the SET button to increment eac...

Page 15: ...6 To get your free copy of the software please register your GammaRAE II at www raesystems com Support ProductRegistration Monitor Set Select Monitor Set to change any of the following parameters Back...

Page 16: ...audible alarm Light only LEDs only or Both off Note You cannot turn off all of the alarm alerts at the same time If you select Both Off when the Vibration alarm is Off Invalid Setting is displayed an...

Page 17: ...are R h or Sv h Press MODE to select your choice and then SET to save your choice and exit Change Date Set the date mm dd yyyy Press MODE to step through month date and year and press SET to incremen...

Page 18: ...perates in 24 hour mode only Press MODE to select Save and press SET to save your time changes Otherwise press MODE again and select Quit Press SET to exit Quit To exit Monitor Set step through the op...

Page 19: ...Settings Data Range Alarm Set Alarm Type Search Search Safety Gamma CF 5 0 1 0 9 9 Gamma Hi 200 cps 0 99999 cps Gamma Low 50 cps 0 99999 cps Datalog Set Clear Data N A N A Change Type Continuous Conti...

Page 20: ...utomatic Manual Off Buzz Lite Both On Both On Light Only Both Off Vibrate On On Off Temp Unit Celsius Fahrenheit Celsius Gamma Unit R h R h Sv h Change Date MM DD YYYY Change Time HH MM SS GammaRAE II...