12. SETTING THE PARAMETERS
User by means of appropriate setting of parameters from menu group
<P4 Readout> can adjust the balance to existing operating conditions
(filter) and one’s expectations (refreshing, autozero, last digit displaying).
Submenu Readout – internal settings
12.1. Filter
For perfect conditions filter can be set as very fast, however if conditions
are bad (vibrations, draughts) filter should be set as slow or as very slow.
Effectiveness of filter operating is different for weighing range. Filter works
slower in time of “reaching” the final result, however stronger when mass
gets to set range of filter operating (parameter range of filter operating
available only in service menu – user do not have access to it).
12.2. Median filter
Task of median filter is to eliminate single big disturbances. By setting
digital value determines speed of filter operating. For zero option filter
operating is off.
12.3. Setting of autozero operating
To ensure accurate balance indications software AUTOZERO function
was introduced. Task of this function is automatic control and correction
of balance zero indication.