4.2. Connectors Arrangement 

 

 

 

– power supply socket 

– RS232 (1) connector 

– weighing platform cable inlet 

– RS232 (2) connector * 

 

*)

 - option. 

 
 

4.3. Pins Overview 

 

 

Pin2 

– RxD 

Pin3 

– TxD 

Pin4 

– 5VDC 

Pin5 

– GND 

RS232 (1) connector, DB9/M (male)

 

 

Pin2 

– RxD 

Pin3 

– TxD 

Pin4 

– 5VDC 

Pin5 

– GND 

RS232 (2) connector, DB9/M (male)

 

 

 

5. UNPACKING AND INSTALLATION 

 

A.  Take the indicator out of the packaging.  
B.  Connect  the  indicator  and  the  platform  and  place  the  set  on  a  flat  and 

even surface. Keep it far away from any sources of heat. 

 
 

6. START-UP AND OPERATION 

 

6.1. Levelling 

 

To level the weighing instrument turn its feet. Keep turning the feet until the air 
bubble takes central position: 

Summary of Contents for PUE C315

Page 1: ......

Page 2: ...2 SEPTEMBER 2020...

Page 3: ...11 3 Taring 14 11 4 Entering Tare Value Manually 14 11 5 Dual Range Devices 15 11 6 Units 15 11 6 1 Start Unit 15 11 6 2 Temporary Unit 16 12 ADJUSTMENT 16 12 1 External Adjustment 17 12 2 User Adjus...

Page 4: ...electing Operation Mode 33 21 2 Setting Sample Mass by Entering Mass of a Single Part 34 21 3 Setting Sample Mass by Determining Mass of a Single Part 34 22 WORKING MODE CONTROL 35 22 1 Local Settings...

Page 5: ...y read this user manual Use the device only as intended B Indicators to be decommissioned must be decommissioned in accordance with valid legal regulations 2 2 Battery Power Supply PUE C315 series ind...

Page 6: ...o polarization Close the lid 3 WARRANTY CONDITIONS A RADWAG feels obliged to repair or exchange all elements that appear to be faulty by production or by construction B Defining defects of unclear ori...

Page 7: ...air is carried out outside RADWAG authorized service point service claims intrusion into mechanical or electronic construction by unauthorized people the device does not bear company protective sticke...

Page 8: ...in2 RxD Pin3 TxD Pin4 5VDC Pin5 GND RS232 2 connector DB9 M male 5 UNPACKING AND INSTALLATION A Take the indicator out of the packaging B Connect the indicator and the platform and place the set on a...

Page 9: ...dication with reading unit next displayed reading unit is conditioned by scale type In case the weighing result is not zero after indication stabilisation press button 6 3 Battery Charge Status The sc...

Page 10: ...tergent soap dishwashing detergent glass cleaner Gently rub the cleaned surface and let it dry Repeat cleaning process if needed In the case of hard to remove contamination e g residues of adhesive ru...

Page 11: ...key press to change the working mode Press to send the weighing result to a printer or computer Press to zero the scale Press to tare the scale Upon pressing keys combination functions of given keys c...

Page 12: ...t Printouts P7 Othr Operation related functions P8 InFo Scale data P9 Unit Units 10 2 Operating the Menu In order to navigate the program menu use the operation panel Press to enter the main menu Pres...

Page 13: ...IGHING Load the weighing pan Read the result when stability marker is displayed Only stable weighing results can be recorded stability marker 11 1 Good Weighing Practice To assure long term operation...

Page 14: ...tion and press key Zero indication and the following pictograms are displayed Net and The weighing device has been tared Upon loading net mass is displayed Taring can be carried out repeatedly within...

Page 15: ...r is displayed in the top left hand corner To return to weighing with the accuracy of the I weighing range unload the weighing pan when zero indication and pictograms are displayed press key II weighi...

Page 16: ...lable units are displayed successively one by one Options in case when the main unit is kg kg kilogram g gram lb pound N Newton unit disabled for verified weighing devices Options in case when the mai...

Page 17: ...be carried out using an external adjustment weight of class F1 Procedure Enter P1 CAL 1 1 CA E submenu text UnLoAd remove weight is displayed Remove the load from the weighing pan and press key Mass...

Page 18: ...Next text UnLoad remove weight is displayed Remove the load from the weighing pan 1 2 CA u submenu is displayed 12 3 Adjustment Report Adjustment report is automatically printed using scale connected...

Page 19: ...ns of the workstation Enter this parameter and set nStAb value if the ambient conditions are unfavourable air drafts vibrations Procedure Enter P2 rEAd 2 3 Enut submenu Press key parameter values are...

Page 20: ...e no Regular tare mode Select this parameter to make the scale overwrite the set selected tare value with the most recently entered one tArF Select this parameter to make the scale store the latest ta...

Page 21: ...are no 1 from tare database is displayed tArE 0 to select a different record press button Select respective entry and press key tare value edit box is displayed Enter tare value to do it press and key...

Page 22: ...ake the last digit always on nEur Select to make the last digit always off uuSt Select to make the last digit on only when the weighing indication is stable Press key to confirm next go to the home sc...

Page 23: ...ared within 0 s 60 s range 0 s is set by default time delay inactive Press key to confirm next go to the home screen In order to establish wireless communication with the scale use RADWAG Connect PC s...

Page 24: ...unication Set the respective value press key to confirm next go to the home screen option 15 1 2 Continuous Transmission Enter 5 1 PC 5 1 2 Cnt submenu Press key parameter values are displayed success...

Page 25: ...er values are displayed successively one by one nonE none port selected rS1 RS232 1 rS2 RS232 2 UUF Wireless communication Set the respective value press key to confirm next go to the home screen opti...

Page 26: ...is automatically generated at the end of each adjustment process Variables list No Name Description 6 1 1 CtP Performed adjustment type 6 1 2 dAt Adjustment date 6 1 3 tin Adjustment time 6 1 4 Idb S...

Page 27: ...the current unit 6 2 6 b Gross weight value in the current unit 6 2 7 CrS Current weighing result net weight in a current unit 6 2 8 CrP The last adjustment report generated in accordance with setting...

Page 28: ...lues are displayed successively one by one where nonE Backlight off 10 Display brightness low limit value in 100 Display brightness high limit value in Set the respective value press key to confirm ne...

Page 29: ...ate and Time Parameter allowing you to set current date and time and to specify date and time format Procedure Enter P7 Othr submenu and change the settings Refer to the below table Parameter Descript...

Page 30: ...Program version 8 3 PStP Settings printout Enter the parameter to send scale settings to printer port all parameters 19 WORKING MODES General Information The weighing device features the following wo...

Page 31: ...1 Acc 3 5 2 Lo Totalizing 3 6 1 Acc 3 6 2 Snn 3 6 3 Int 3 6 4 Lo Animal weighing 3 7 1 Acc 3 7 3 Lo The table presents special function number and name for each of the working modes Remaining specifi...

Page 32: ...n min to set the interval go to Int parameter Set the respective value press key to confirm next go to the home screen 19 2 3 Automatic Printout Time Interval Parameter enabling you to set frequency o...

Page 33: ...e same mass which determination is done on the basis of mass of sample piece single part and where the sample piece mass single part mass is determined using the weighing device 21 1 Local Settings To...

Page 34: ...message Err Hi is displayed 21 3 Setting Sample Mass by Determining Mass of a Single Part Enter 3 2 PcS 3 2 2 UUt submenu set S_S value Enter PcS working mode parts counting blinking value of sample q...

Page 35: ...local settings enter 3 3 HiLo submenu 3 3 1 Acc Working mode accessibility For detailed description read section 19 2 1 3 3 2 Snn Save mode For detailed description read section 19 2 2 3 3 3 Int Time...

Page 36: ...tings enter 3 4 dEu submenu 3 4 1 Acc Working mode accessibility For detailed description read section 19 2 1 3 4 2 UUt Operation mode For detailed description read section 23 1 1 3 4 3 Snn Save mode...

Page 37: ...SEt_Ut is displayed for 1 s next window for declaring mass value of the reference sample Enter the value and press key to confirm The home screen is displayed automatically with 0 000 value If the val...

Page 38: ...llows you to sum up to 30 weighings ingredients maximum within one process 25 1 Local Settings To go to local settings enter 3 6 Add submenu 3 6 1 Acc Working mode accessibility For detailed descripti...

Page 39: ...nted on a scale connected printer Report example In order to print the report once again press key To exit report printout mode press key As a result the home screen of Add working mode is displayed a...

Page 40: ...tion Enter AnLS working mode Animal Weighing First text tinnE is displayed for 1 s next window for setting duration in seconds of animal weighing Press key parameter values are displayed successively...

Page 41: ...28 TECHNICAL SPECIFICATIONS Housing Plastic Ingress protection IP43 Operating temperature 10 C do 40 C Display LCD backlit Power supply 100 240 VAC 50 60 Hz 12 24 VDC Battery power supply 6 x NiMH AA...

Page 42: ...start up message LH is displayed Weighing pan loaded during the start up Unload the weighing pan Zero indication is displayed Communication with the computer not established Incorrect computer port se...

Page 43: ...eyond zero range E r r 3 Value beyond tare range E r r 4 Adjustment weight or start mass out of range 1 for adjustment weight 10 for start mass E r r 5 Battery error Battery is damaged E r r 8 Time of...

Page 44: ...44...