background image

 

- 62 - 

 

18.4.14.  Unlock the scale keyboard 

 

Syntax: 

K0 CR LF

 

 

Possible answers: 

K0_OK CR LF 

– command in progress 

 
 
 

18.4.15.  Give serial number 

 

Syntax:  

NB CR LF

 

 

Possible answers: 

 

NB_A_”Factory number” CR LF 

- command comprehended, scale serial  
   number is given in return  

NB_I CR LF 

- command comprehended but cannot be  
  executed 

 

”Factory number”

 – parameter specifying scales serial number, it is 

returned in between inverted comas. 

 

Example: 

NB CR LF

 – command from a computer 

NB_A_”123456” CR LF

  – scales serial number - 123456 

 
 
 

18.4.16.  Send all implemented commands 

 

Syntax: 

PC CR LF

 

 

Possible answers:  

 

PC_->_Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,K1,K0,OT,UT,NB,PC  
 

– command executed, the indicator have sent all the implemented 

commands. 

 

 
 

18.5. Manual printouts / automatic printouts 

 

Users can general manual or automatic printouts from the scale.  

 

  Manual printouts can be performed after loading the pan  

and stabilizing indication by pressing 

.  

Summary of Contents for PUE C/31

Page 1: ...PUE C 31 WEIGHING INDICATOR PUE C 31 Indicator PU C 31H Indicator PUE C 31H Z Indicator USER MANUAL ITKU 17 02 08 16 EN...

Page 2: ...2 AUGUST 2016 The drawings photos and graphics used are for illustrative purposes only...

Page 3: ...emporarily selected unit 18 11 MAIN PARAMETERS 19 11 1 Setting a filtering level 19 11 2 Median filter 20 11 3 Autozero function 21 11 4 Tare function 22 12 RS 232 PARAMETERS 23 12 1 Printout type 23...

Page 4: ...interfaces 55 18 3 Respond message format 56 18 4 Command s description 56 18 4 1 Zeroing 56 18 4 2 Tarring 57 18 4 3 Get tare value 57 18 4 4 Set tare value 57 18 4 5 Send the stable result in basic...

Page 5: ...RS 232 designed printouts designation minimum mass for function operating counting pieces mass control percentage deviation from standard mass latch of maximum scale indication automatic tare memory o...

Page 6: ...pplied from NiMH batteries nickel metal hydrogen with rated voltage of 1 2V size R6 and capacities from 1800 to 2800mAh charged while connected to mains without stopping operation PUE C 31H and PUE C...

Page 7: ...n use to power the device R6 size standard non rechargible batteries If you want to use normal batteries instead of rechargeable ones proceed as follows Before installing non rechargeable batteries tu...

Page 8: ...H Z weighing indicators stainless steel housing the worn out accumulator can be exchanged to a new one by the authorized service of the manufacturer 2 3 Operation in a strong electrostatic field If th...

Page 9: ...s D Warranty does not cover Mechanical failures caused by inappropriate maintenance of the device or failures of thermal or chemical origin or caused by atmospheric discharge overvoltage in mains or o...

Page 10: ...dimensions of PUE C 31 Main dimensions of PUE C 31H PUE C 31H Z 5 GETTING STARTED After connecting platform to a PUE C 31 indicator put the platform on a flat stable ground far away from sources of h...

Page 11: ...e test completion Then you will see zero indication and pictograms zero indication stable result kg weight unit If the indication is not zero press zero key 6 KEYPAD 7 KEYS FUNCTIONS Switching on off...

Page 12: ...h of the max measurement 10 Add Totalizing 11 AnLS Weighing animals 12 tArE Memory of 9 tare values 13 Indication in autozero zone indication exact zero 14 Stable result ready to read 15 PCS Operation...

Page 13: ...YES P 1 3 tArA no P 1 4 Fnnd no P2 Prnt P2 1 Pr_n StAb P2 2 S_Lo P2 3 bAud 9600 P2 4 S_rS 8d1SnP P3 Unit P3 1 StUn kg P4 Func P4 1 FFun ALL P4 2 Funi no P4 3 PcS no P4 4 HiLo no P4 5 PrcA no P4 6 Prcb...

Page 14: ...eter or changing the value of a selected parameter Entering the selected submenu or activating a parameter for changes Confirmation enter Leaving without changes or reaching a higher level of the menu...

Page 15: ...ro will be indicated After placing a load on the weight pan net mass will be shown Tarring is possible within the whole range of the scale After unloading the pan the display shows the tarred value wi...

Page 16: ...ort audible signal will be emitted 2 Users can also enter up to 9 tare values to the scale memory see 14 10 of his manual 10 3 Zeroing To ZERO the scale press The scale will display zero and following...

Page 17: ...n the top left corner of the display Then weighings is done with the accuracy of the II range to the moment of returning to zero autozero range where the scale switches back to the I range 10 5 Select...

Page 18: ...for verified scales lb is not accessible After you select the unit press the scale returns to Return to weighing according to chapter 9 2 2 Notice After turning on the scale always sets the basic unit...

Page 19: ...rs can select following units g ct lb lb is not accessible for verified scales 11 MAIN PARAMETERS Users can adjust the scale to external ambient conditions filtering level or particular needs autozero...

Page 20: ...vel is the longer stabilization time is needed Return to weighing See 9 2 2 11 2 Median filter This filter eliminates short changes impulses of measure signal e g shocks Procedure Enter the submenu P1...

Page 21: ...declared value of autozero range so the scale will be automatically zeroed and the pictograms and will be displayed When AUTOZERO is disabled zero is not corrected automatically However in particular...

Page 22: ...c tare tArA no automatic tare function off user can turn on operating of automatic tare F6 AtAr till unplugging the balance from mains tArA tArF tare memory function stores last value of tare in balan...

Page 23: ...even odd There are four ways of sending data via RS232 interface Manually after pressing Automatically after stabilizing the indication over LO threshold Continuously after it is activated in paramet...

Page 24: ...necessary while working with automatic tare or automatic operation or weighing animals Automatic tarring will not be applied until the indication gross is lower than the value inscribed in S_Lo parame...

Page 25: ...25 Return to weighing see 9 2 2 12 3 Baud rate Procedure Enter the submenu P2 Prnt and then...

Page 26: ...s no parity control 7d1SEP 7 data bits 1 stop bit EVEN parity control 7d1SoP 7 data bits 1 stop bit ODD parity control 8d1SnP 8 data bits 1 stop bit no parity control 8d2SnP 8 data bits 2 stop bits no...

Page 27: ...unction Program recognises the way the scale is supplied mains battery and automatically selects the way of operating on the backlight bl for mains blbt for batteries or rechargeable battery pack 13 1...

Page 28: ...light for supplying from batteries The user can change the intensity of backlight from 0 to 100 The lower the intensity is the longer the scale operates without recharging or exchanging batteries When...

Page 29: ...ower The scale is switched off automatically when function t1 YES no weighing appears in 5 minutes no changes on the display In case when this function disrupts the operation e g long time weighing pr...

Page 30: ...ery voltage level check While supplying from batteries too low level of voltage is measured by software the pictogram is displayed It means that charging or exchanging batteries is required 13 4 1 Che...

Page 31: ...omatically switch off to protect the batteries from distructable discharging Charging is signalled by blinking period about 2 seconds on the display 13 4 3 Accumulator charging option This function al...

Page 32: ...geable batteries NiMH R6 AA and power adapter Procedure Enter the submenu P5 othr and then CHr6 YES enabled CHr6 no disabled Return to weighing See 9 2 2 13 4 4 Formatting rechargeable battery packs E...

Page 33: ...moment of self powering down 5 Repeat the process of charging starting from point 1 Notice They reach their optima capacity after three cycles of full charging and discharging 14 WORK MODES 14 1 Sett...

Page 34: ...function enables user to set if after pressing key all operating modes will be accessible ALL or only one from the list chosen and used by operator Procedure Enter the submenu P4 Func and then After...

Page 35: ...e in order to tare a container value Notice 1 Counting pieces does not work together with other scale functions 2 The counting pieces function is not saved as a default start function so it is not rem...

Page 36: ...er the required sample quantity where selection of digit position setting the digit Confirm the value by pressing You will see LoAd on the display and then If weighing is performed in a container put...

Page 37: ...ng unit mass adjustment Single piece mass should not be less than 5 divisions 3 If a single piece mass is lower than a reading interval d the display will show the Err5 message see ch 19 Error message...

Page 38: ...x The inscribed value confirm by pressing the program will automatically go to the main window During setting threshold values following cases take place Notice If a user erroneously enters a value of...

Page 39: ...or entered to the scale memory by an user PrcB function 14 5 1 Standard mass determined by its weighing Procedure Enter to PrcA function You will see LoAd on the display and then place an load on the...

Page 40: ...ogram goes to the weight display window Using and set standard mass where digit selection digit setting Confirm the entered value by pressing You will see the indication equal to 0 000 From this momen...

Page 41: ...hen indication is stable automatic tarring of the container mass will be performed Net marker will appear in the upper part of the display place a sample into the package display will indicate net mas...

Page 42: ...ce to the weighing pan The display of scale will latch the maximum value of the force remove loads from the pan Before the next measurement press the key Return to weighing Press the key twice 14 8 To...

Page 43: ...g to ch 14 8 1 Put the first load on the pan If the weighing procedure is performed in a container put the container on the pan first and tare it Then put the first load on the pan and confirm it by p...

Page 44: ...sing one more time with loaded pan you will see the unLoAd message Unload the pan the scale will return to ZERO and the letter P in the left part of the display will appear The scale is ready for the...

Page 45: ...ll see Before leaving the Add function it is possible to print out subsequent weighings and the sum of weighings on the connected printer press to print press to cancel The following message will appe...

Page 46: ...AnLS function The tinnE message appears on the display for 1s and then the program goes to the window of setting the duration time in seconds of the animal weighing process Confirm the selected value...

Page 47: ...tched on the display and additionally the OK pictogram is shown in the upper part of the display You can start the procedure of weighing animals again by pressing After removing the animal from the pl...

Page 48: ...alue from the selection of tare values tArE 0 press to chose different values After selecting the right position press and you will see an editing field Enter the selected tare value to the scale memo...

Page 49: ...e tare value on the display preceded by the sign and the Net pictogram Caution A tare value from the memory is not remembered after powering off and on the scale 15 USER CALIBRATION Only for non verif...

Page 50: ...ss is adjusted during this period of time After that a mass of calibration weight is shown e g 3 000kg Put a weight of the displayed mass value on the pan and press The calibration process will start...

Page 51: ...hanges that have been made Caution If the calibration process span adjustment lasts longer than 15 the Err8 message will be displayed and short audible signal will be emitted Press to perform calibrat...

Page 52: ...rt mass adjustment can be terminated by pressing which is signalled on the display Return to weighing performing the procedure of saving parameters Caution If the start mass adjustment lasts longer th...

Page 53: ...ding on setting of REPL parameter printout will be automatic or manual Cable diagrams Scale printer cable diagram for plastic casing 17 COOPERATION WITH COMPUTER Sending weighing results to the comput...

Page 54: ...nt information from the scale display The program allows to configure easily e g design printouts edit parameters A precise description is issued in the Help file that accompanies the program Cable di...

Page 55: ...e operation e g requesting weighing results display control 18 2 A set of commands for RS interfaces Commands Description of commands Z Zeroing T Tarring OT Get tare UT Set tare S Send the stable resu...

Page 56: ...mmand comprehended but the indication below the XX _ OK CR LF Command done ES_CR LF Command not comprehended XX _ E CR LF error while executing command time limit for stable result exceeded limit time...

Page 57: ...mprehended but cannot be executed 18 4 3 Get tare value Syntax OT CR LF Possible answers OT_TARA CR LF command executed Frame format 1 2 3 4 5 6 7 15 16 17 18 19 20 21 T O space stability space tare s...

Page 58: ...FRAME command accepted and in progress mass value in basic unit is returned Frame format 1 2 3 4 5 6 7 15 16 17 18 19 20 21 S space stability space sign mass space unit CR LF Example S CR LF computer...

Page 59: ...epted and in progress timeout while waiting for stable results SU_I CR LF command comprehended but cannot be executed SU_A CR LF MASS FRAME command accepted and in progress mass value in current unit...

Page 60: ...eturned 18 4 9 Switch on continuous transmission in basic unit Syntax C1 CR LF Possible answers C1_I CR LF command comprehended but cannot be executed C1_A CR LF MASS FRAME command comprehended and in...

Page 61: ...5 6 7 15 16 17 18 19 20 21 S U I stability space sign mass space unit CR LF 18 4 12 Switch off continuous transmission in current unit Syntax CU0 CR LF Possible answers CU0_I CR LF command comprehende...

Page 62: ...cales serial number it is returned in between inverted comas Example NB CR LF command from a computer NB_A_ 123456 CR LF scales serial number 123456 18 4 16 Send all implemented commands Syntax PC CR...

Page 63: ...ve values mass 9 characters justified to the right unit 3 characters justified to the left command 3 characters justified to the left Example 1 _ _ _ _ _ _ 1 8 3 2 0 _ g _ _ CR LF the printout generat...

Page 64: ...racters justified to the right unit 3 characters justified to the left command 3 characters justified to the left The frame format sent by the indicator in case of setting P2 Prnt to Cntb 1 2 3 4 5 6...

Page 65: ...5 to 15 of start mass 5 FULL Display range overflow in totalizing Notice 1 Errors Err2 Err3 Err4 Err5 Err8 Err9 null that appear on the display are also signalled by a short beep sound about 1 sec 2 E...

Page 66: ...crement of signal 19 2 mV Max Voltage per 1verificated division 3 2 V Min Voltage per 1verificated division 1 0 V Max Temperature of work 40 C Min Temperature of work 10 C Min Impedance of load cell 8...

Page 67: ...IPMENT Accessories Computer cable for PUE C 31 P0108 Computer cable for PUE C 31H PUE C 31H Z P0259 EPSON printer cable for PUE C 31 P0151 EPSON printer cable for PUE C 31H PUE C 31H Z P0261 Power cor...

Page 68: ...rter for PUE C 31 KR 04 A rack for PUE C 31 PUE C 31H or PUE C 31H Z indicator Handle for measuring indicator in plastic version Computer programs EDYTOR WAG computer program RAD KEY computer program...

Page 69: ...69...