- 44 -
15.3. Levelling
The balance is equipped with an AutoLEVEL System which enables monitoring of the
balance level. Level monitoring is performed uninterruptedly in the course of balance
operation. Level status is signalled with a respective pictogram, displayed in the upper right
corner of the work panel. The system monitors balance level state and on detecting any
change the position of level marker is displayed on the screen and/or respective alarm is
activated. Balance level setting proceeds.
Levelling procedure:
•
Press level status
pictogram located on the top of the display.
•
Control panel of level function is displayed . Along with the level indicator, balance
view is presented.
•
Level the balance by turning its feet left/right, follow the blinking pictograms, <
> –
press to make level marker move toward the centre of the circle.
•
When the marker takes central position, its colour turns from red to green – the
balance is correctly levelled.
Caution: The balance is equipped with Automatic Level Control mechanism. For detailed
information regarding the mechanism operation refer to section 34.8 of this user manual.
15.4. Zeroing
To zero mass indication press
key. Zero indication and the following pictograms are
displayed:
and
. Zeroing operation means determining new zero point, recognized by
the weighing instrument as precise zero. The instrument can be zeroed only when the
indication is stable.
CAUTION:
Indication can be zeroed only within ±2% range of maximum capacity. If the zeroed value is
above ±2% of the maximum capacity, then the software indicates a respective Err2 error
message.
15.5. Taring
To determine net weight value, load the weighing pan with a packaging, wait for a stable
indication and press
key. Zero indication and the following pictograms are displayed:
Net
and
. Upon unloading the weighing pan, the sum of tared masses with minus sign is
displayed. You can assign tare value to a particular product stored in the database. In such
case the assigned tare value is automatically acquired upon selection of the given product.
Summary of Contents for MAS 1. Y
Page 1: ...WEIGHING MODULES MAS Y MPS Y MAS 1 Y USER MANUAL IMMU 25 01 08 18 EN www radwag com...
Page 9: ...9 1 GENERAL INFORMATION 1 1 Dimensions MAS Module MPS Module...
Page 10: ...10 MAS 1 Module...
Page 200: ...200...