12

 - 

 

13. Select „Operating Mode” tab. 

 

 

 

By default „Operating mode” is set to „TCP Server Mode”. 
By default „Local TCP Port” is set to 4001. 

 
14. Press „OK” button to confirm changes. 
15. Exit 

Network Enabler Administrator

 PC software. 

 
Setup parameters may vary depending on individual needs. 
 

Network Enabler Administrator 

PC software is to be downloaded from www.radwag.com. 

Summary of Contents for KR-04 Series

Page 1: ...KR 04 RS232 Ethernet converter ITKU 92 01 07 16 EN www radwag com USER MANUAL ...

Page 2: ... 2 JULY 2016 ...

Page 3: ... 3 CONTENTS 1 INTENDED USE 4 2 TECHNICAL SPECIFICATIONS 4 3 DESIGN 4 3 1 External view 5 3 2 Connectors 7 4 SETUP 9 ...

Page 4: ...HTTP SNMP Operating modes TCP Server TCP Client UDP Operating temperature 0 40 C 3 DESIGN Converter types KR 04 1 plastic housing RS232 cable with DB9 F connector Intended for scales equipped with PUE C31 plastic housing laboratory balances and other devices featuring RS232 interface with DB9 M connector KR 04 1 converter is equipped with Ethernet interface of RJ45 type KR 04 2 stainless steel hou...

Page 5: ... 5 3 1 External view KR 04 1 converter 1 Ethernet RJ45 port 2 RS232 cable with DB9 F connector 3 power supplier Dimensions of KR 04 1 converter plastic housing ...

Page 6: ... 6 KR 04 2 KR 04 3 converter Dimensions of KR 04 2 KR 04 3 converter stainless steel housing ...

Page 7: ... RS232 cable cable with RJ45 connector KR 04 2 KR 04 3 interfaces 1 power cord cable gland 2 RS232 cable gland cable with a hermetic connector 3 Ethernet cable gland KR 04 2 Ethernet hermetic socket KR 04 3 Ethernet connector pin description Pin1 RD Pin2 TD Pin3 RD Pin4 TD Ethernet connector KR 04 3 ...

Page 8: ...KR 04 3 1 Ethernet hermetic connector 2 shielded twisted pair wire 4x2xAWG26 3 RJ45 connector Ethernet cable KR 04 2 1 shielded twisted pair wire 4x2xAWG26 2 RJ45 connector 3 ring terminal 4 cable LgY 300 500V 0 35mm2 ...

Page 9: ...k Enabler Administrator PC software 2 Establish communication between the computer and the converter via LAN use crossed cable terminated with RJ45 connectors on both sides For KR 04 3 converter use RADWAG manufactured P0198 cable and a switch 3 Connect the converter to the mains 4 Run Network Enabler Administrator PC software 5 Click Zoom icon the software will search for network connected conver...

Page 10: ...work is displayed 7 Select the respective converter double click using left mouse button 8 Go to Configuration window select Network tab 9 Tick Modify box by field that is to be changed 10 Enter all necessary data 11 Select Serial tab RS232 setup ...

Page 11: ...2 Tick Modify box and select Port that is to be modified double click using left mouse button Default settings Flow Control NONE FIFO Enable Interface RS232 Baud rate 9600 Parity NONE Data bits 8 Stop Bits 1 ...

Page 12: ...rver Mode By default Local TCP Port is set to 4001 14 Press OK button to confirm changes 15 Exit Network Enabler Administrator PC software Setup parameters may vary depending on individual needs Network Enabler Administrator PC software is to be downloaded from www radwag com ...

Page 13: ... 13 ...

Page 14: ... 14 ...

Reviews: