- 90 -
value. Bottom line of the display, if <BARGRAPH> option is enabled, shall
demonstrate weight indication, in a graphic form, in relation to set
thresholds.
<Min>: mass lower than the value of the low threshold
<Ok>: mass contained within thresholds
<Max>: mass higher than the value of the high threshold
17.3.2 Return to weighing mode
17.4
Dosing
Dosing mode comprises process of sample weighing, wherein the said sample
weighing operation is performed until target mass is reached. The target mass is
defined along with dosing tolerance value. The tolerance value is set as
a percentage of the target mass.
Example:
Target weight = 100.000g
Tolerance = 2,5% (2,5% of 100g, which amounts to 2,5g)