12
7.1.2. CY10.xx.K Series Scales
A. Remove the scale from the factory box.
B. Place the device in the place of use, on the even and hard flooring, away
from sources of heat.
C. Pull transportation protection out and put the weighing pan on:
7.2. Scale Leveling
Turn adjustment feet to level the scale. Leveling is correct if the air bubble is in
the center of the level that has been placed on the foundation:
7.3. Connection to Network
The device can be plugged in only with the use of the original feeder cable
supplied. The rated voltage of the feeder cable (given in dedicated rating plate)
should correspond to the mains rated voltage.
Procedure:
Plug the feeder cable into the mains socket and put the feeder cable plug
into the port at the back of the weighing indicator housing.
The operating system and RADWAG software loading procedure will
initiate in a moment. While the program is being launched, a signal LED
light and LED lights at the lower frontal part of the indicator will flash.
Once the start-up procedure has ended, the home page will be displayed.
The weighing indicator starts with the operator being unlogged (no user).
To start working, log in (logging procedure has been described below).
Switch the scale on without any load
– with the empty
weighing pan. As per EN 45501 standard, in verified scales
the mass value of less than -20e must not be displayed.
Therefore when the result drops below this value,
the home page will display <Lo mass>. This being the
case, zero the scale by pressing
.