- 37 -
•
adjustment triggered by elapsing time. It is possible to declare time
intervals which are a criteria for balance adjustment. Accessible
settings are: adjustment after 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12 hours
since last carried out adjustment procedure,
CAUTION
Declaring time interval for balance adjustment is available only in
non-verified balances. In case of verified balances the time interval
between following adjustments is set for 3 hours.
12.2.
Menu adjustments settings
P1 CAL
P1.1
iCAL
|
P1.2
ECAL
|
P1.3
tCAL
|
P1.4
ACAL
|
both
P1.5
CALt
|
1 h
P1.6
CALr
|
YES
•
P1.1 iCAL – internal adjustment
Start of internal adjustment process. Adjustment is conducted automatically,
and no operator assistance is required. If weighing pan of the balance is
loaded, balance will display a command to unload the weighing pan. If
weighing pan is loaded with relativaly small mass (up to 4 % of balance max
capacity) than adjustment process will be carried out automatically, but
measuring results may differ by the amount that was present on balance
weighing pan.