![RADWAG AS R2 PLUS Startup Manual Download Page 22](http://html1.mh-extra.com/html/radwag/as-r2-plus/as-r2-plus_startup-manual_3376238022.webp)
11.3. Weighing chamber elements disassembly
Prior disassembling weighing chamber elements make sure the balance is turned off and unplugged
from the mains. When disassembling the elements, use tweezers that come standard with the
instrument.
1
2
4
3
Press the moldings on the housing and lift the movable element up.
Press the moldings
on the housing and lift
the movable element up.
Now you can take out the front glass.
Now you can take out
the upper glass.
5
6
Take out the back glass.
The housing has recesses
to facilitate removing the glass.
Slide out the door.
After cleaning the weighing
chamber elements, they must be
assembled in the reverse order.