4. Sample measurements
ABL800 FLEX Operator's Manual
General information
Measuring
modes with
FLEXQ
With the FLEXQ module installed, it is possible to make the measuring process
automatic (i.e. just place up to three samples in the sampler tray and go), or to
place a single sampler in the sampler tray and perform a single measurement.
The following rules apply to the
batch mode
:
•
The analyzer is connected to the RADIANCE system
•
All information is entered on the RADIANCE system via FLEXLINK
•
Automatic data request is based on Sampler ID
•
Measuring mode will always be the largest syringe mode, and it will be
marked "Flex"
•
Default layout is always used.
The following rules apply to a
single measurement
:
Parameter profile received from the
RADIANCE system
Parameter profile not received from
the RADIANCE system
•
Measuring modes will be those
syringe modes that can provide the
requested parameter profile; these
modes will be marked "Flex"
•
The measuring modes are those that
can be selected on the analyzer (see
Available modes and parameters
below)
•
Default layout is always used.
•
The layout will be the one assigned
to the selected measuring mode.
Available modes
and parameters
The modes and the measured parameters available on the ABL800 FLEX
analyzers are listed below.
Analyzer
Measuring modes
Measured parameters
805
Syringe – S165
μ
L
pH,
p
CO
2
,
p
O
2
,
c
K
+
,
c
Na
+
,
c
Ca
2+
,
c
Cl
–
,
c
Glu,
c
Lac
Syringe – S95
μ
L
Capillary – C95
μ
L
pH,
p
CO
2
,
p
O
2
,
c
K
+
,
c
Na
+
,
c
Ca
2+
,
c
Cl
–
,
c
Glu,
c
Lac
Syringe – S85
μ
L pH+BG
Capillary – C55
μ
L pH+BG
pH,
p
CO
2
,
p
O
2
Capillary – C35
μ
L MET
c
Glu,
c
Lac
Capillary – FLEXMODE
See below
Syringe – Pleura pH*
pH
Other fluids (any mode)
Depends on the selected
mode
Expired
air
p
CO
2
,
p
O
2
* See page 4-5
Continued on next page
4-2
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...