TitraLab 870 Reference Manual
Page 247
Temperature
Probe/ Fixed at
25°C/Entered
Probe
The sample or standard temperature can be measured using a tem-
perature sensor connected to the
Temp
socket of the Titration
Manager. Select a temperature sensor in the Method parameters
menu.
Fixed at 25°C
If measurements are to be performed at a constant temperature of
25°C.
Note:
No temperature correction is available for conductivity measurements.
Entered
If the temperature is to be entered manually.
Select in:
Edit method menu
Edit reagent menu if Titre = Calibrate
Edit electrode menu if Calibrate request = Yes
Temperature
sensor ID
Name of the temperature sensor or name of the electrode with a built-
in temperature sensor.
Enter in:
Edit Electrode > Calibration parameters menu
Edit Method > Method parameters
Edit reagent > Reagent parameters
Range available:
16 characters
Test amount
Sample quantity to be placed into the beaker.
Enter in:
Menu Beaker (Menu Sequence/Sample stack then press
1
Sample
stack).
Refer to "Sample stack", page 225
.
Range available:
0,001 to 100000 (unit =
Sample Unit
)
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...