TitraLab 870 Reference Manual
Page 227
Select reagent
Routine user
Is able to select a reagent to check the parameters and/or start a
calibration.
1.
Select the method or the sequence using the reagent.
2.
Press
1
to start a calibration or
3
to check the reagent parame-
ters. In both cases, the available reagents are those specified in
the actual working method or sequence.
Supervisor
From the Reagent window.
1.
Press
4
.
2.
In the ID field, select the reagent from the User list.
Select
sequence
To select a sequence (sample changer not in use):
1.
Select
Working mode = Sequence
in the Main window.
2.
Press
2
Sequence/Sample stack.
To select a sequence to be run using a Sample Changer:
1.
In the Configuration menu, define the model of sample changer
used.
Refer to "Sample changer", page 221
.
2.
Select
Working mode = SAC Sequence
in the Main
window.
3.
Press
2
Sequence/Sample stack.
Refer to "Sequence/SAC sequence", page 228
Sensitivity
Measure of the electrode condition. For ideal electrode chains the
sensitivity is 100%. However, it is generally lower. It is expressed as a
percentage of the theoretical slope (59.16 mV/pH) of the curve at
25°C and is determined during a calibration on at least 2 points.
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...