Titra
Lab
®
TitraLab 870 Reference Manual
Page 108
Electrode
calibration
(conductivity)
1.
Select the method which uses the conductivity cell to be
calibrated.
2.
see "Electrode connection", page 114
.
3.
Press
1
Calibrate electrodes in the Electrode window.
4.
Select the conductivity cell from the list.
5.
Press
1
to Run, and follow the messages on the display.
If you are not using a temperature probe and want to get
accurate measurements, it is recommended to thermostat
your standard beakers at the temperature you have entered
(or at 25°C).
Pay attention to the temperature range of the standard used:
see "Standard (conductivity standard)", page 235
6.
After stabilisation or at the end of the maximum stabilisation time,
the TIM870 calculates the standard conductivity at the measured
or entered temperature. Then, the instrument calculates and
displays the cell constant.
Refer to "Result accepted (Yes/No)", page 205
.
7.
The cell constant (mean of all cycle results accepted) is saved
with the electrode.
To consult the calibration results,
The TIM870 displays the conductance
measured. The displayed temperature is
the temperature measured, entered or is
equal to 25°C according to the
calibration method programmed.
Accept or reject the result then start a
new calibration cycle (new beaker of the
same standard) or end the calibration.
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...