TitraLab 870 Reference Manual
Page 114
Electrode
connection
Proceed as follows to connect/install electrodes and temperature
sensors:
1.
In the Electrode window, press
2
then
1
Connect electrode.
2.
Enter serial number.
3.
Install electrodes on the Titration Manager (TIM), ABU52 or
sample changer (SAC80, SAC90, SAC850 or SAC950).
4.
Connect electrodes to the rear panel socket of the Titration
Manager (TIM) or ABU52 (ABU1 or ABU2),
. See figure and table below. For example: pHC2001 to
address TIM/E1.
5.
Press
1
to confirm.
Figure 20: Electrode sockets
Table 3: Connecting electrodes
If the current method in use requires an electrode different to
the one already connected, the Titration Manager will prompt
you to disconnect the electrode before connecting the new
one,
see "Disconnect electrodes", page 97
.
Refer to "Electrode connection - Important", page 115
P
t
-P
t
R
ef
E1
E2
T
emp
GND
EC
Socket
Type of electrode
REF
Single reference
TEMP
Temperature
GND
Ground metal for cell grounding only
Pt-Pt
Double metal
E1/E2
Indicating (pH, Metal/Redox, ISE) single or combined
EC
Conductivity cell w/o temperature sensor
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...