4.2 Get the room temperature
4.2.1 Setting “Real time clock” first to enable this function. After power off, room temperature will be shown on the screen with
icon.
4.2.2 For room temperature, the thermometer should be placed on the table and avoid the direct sunshine or air conditioner flow on.
4.2.3 The room temperature can be reference after 15min later.
5. After measurement:
5.1 Power off: Device will automatically shut off if left idle for more than 1 minute to extend battery life.
The LCD will only show the date, time and room temperature.
5.2 Clean the probe after each use to ensure an accurate reading and avoid cross contamination. (See the section of Care and Cleaning for details.)
5.3 Remember to put on the Probe Cap when not in use.
Use of the Stopwatch
1.
Press and hold the “ON/MEM” button, and press the “SCAN” button twice for “stopwatch” mode.
2.
For start or stop counting, press the “SCAN” button; for reset counting, press the “ON/MEM” button.
3.
Press and hold the “ON/MEM” button, and press the “SCAN” button to leave this mode.
4.
How to measure the heartbeat? If you count out the number of beating is 18 in 15sec, then you can calculate the heartbeat of patient is 72 times in 1min.
Important Notes
Care and cleaning
T
he probe is the most delicate part of the thermometer. Use with care when cleaning the lens to avoid damage.
a. After the measurement, please use the cotton swab with the Alcohol (70% concentration) to clean the lens
( on the inside of the probe).
b. Allow the probe to fully dry for at least 1 minute.
c. Storage temperature range: It should be stored at room temperature between -20~+50°C , RH
≦
85%
d. Keep the unit dry and away from any liquids and direct sunlight.
e. The Probe should not be submerged into liquids.
* Note: Please check the device if damaged once it falls. If you can’t make sure of it ,please send the complete device to
the nearest retailer for recalibration.
Holding the thermometer too long may cause a higher ambient temperature reading of the probe.
This could make the body temperature measurement lower than usual.
Battery replacement
Keep the battery away from children.
(This device is supplied with one lithium cell (CR2032x1).)
○
1
Open the battery cover: Insert a pointed object, such as a pen, into battery cover pick
hole. At the same time, use thumb to push battery cover out.
○
2
Hold the device and flip the battery out with a small screwdriver
○
3
Insert the new battery under the metal hook on the left side and press the right side of
the battery down until you hear a click
* The positive (+) side up and the negative (-) side pointed down.
○
4
Replace the battery cover
Trouble shooting:
Error Message
Problem
Solution
Error 5~9, the system is not functioning properly.
Unload the battery, wait for 1 minute and repower it. If the message
reappears, contact the retailer for service.
Measurement before device stabilization.
Wait until all the icons stops flashing.
The device showing a rapid ambient temperature change.
The ambient temperature is not within the range between 10
o
C and 40
o
C (50
o
F ~104
o
F).
Allow the thermometer to rest in a room for at least 30 minutes at room
temperature: 10
o
C and 40
o
C (50
o
F ~104
o
F).
Temperature taken is higher than +42.2
o
C (108°F)
Temperature taken is lower than +34
o
C (93.2°F)
Please select the target within specifications. If a malfunction still exists,
please contact the nearest retailer.
Device cannot be powered on to the ready stage.
Change with a new battery.
Warranty
Warranty :12 months
Manufacture Date
:
as the serial number( please open the battery cover, it is shown on the inside of the device.)
Ex.SN:E912A000001, the first “E” is External, the second number “9” is the manufacture year 2009, the third and the fourth number “12” is the manufacture month, the
others is the serial number.
Note: Suggest that the device is performed the re-test for accuracy after 3 years. Please send the complete device to the dealers or nearest service address.
However, if this device is used according to the operation instruction, periodic re-calibration is not required.
Please read the instructions for use
BF type applied part
Radiant Innovation Inc. Http://www.radiantek.com.tw
Add: 1F, No.3, Industrial East 9
th
Road, Science-Based Industrial Park,
HsinChu, Taiwan 300.
Medical Technology Promedt Consulting GmbH
Add: Altenhofstrasse 80, D-66386 St. Ingbert, Germany
Ref No.
:
052012