background image

 
 
                                                                                       

Radiant Innovation 

5

                       
3. 

Measuring the temperature:

 Aim the thermometer on the target and press the 

“Meas.” key to take the temperature.   

 

  

 

 
 

Note: 

Since the viewing angle of TN1 is large, please keep the thermometer as near as 

1~2cm to the target to get the accurate target temperature. The information of viewing 
angle and area is shown on the sticker attached on the rear side of the thermometer. 
 
 

Warning: 

 

This Infrared Thermometer (TN1) is calibrated for Emissivity=0.95 only.  Most of 

Non-metal materials have emissivity near to 0.95. But the infrared emissivity of 

normal metal is much lower. TN1 may give error reading for such materials. For 

example, a 100

 °

C boiling pot (aluminum, emissivity~0.3) may show only 50

 °

C, 

when you measure the temperature of the aluminum! 

 
 

 

4. 

Power off:

 Device will automatically shut off if left idle for more than 15sec to 

extend battery life. 

   

 

OPERATION INSTRUCTIONS

Summary of Contents for TempFairy TN1

Page 1: ...Instruction Manual...

Page 2: ...and uses of the TN1 series thermometer 2 Parts Description 3 How to operate 4 Unit Switch between Fahrenheit and Celsius 6 Mode Functions 6 Cleaning and Storage 8 Changing the Battery 8 Error Message...

Page 3: ...temperatures from 22 C to 110 C 7 6 F to 230 F TN1 also has mode adjusting function to scan the minimum or maximum temperature The thermometer also provides a LOCK mode to monitor the temperature of...

Page 4: ...on 3 Parts Description 1 LCD Display 2 Battery Cover 3 Measure Key 4 Mode Key 5 Sensor 6 Stand 7 Key Ring Hole PARTS DESCRIPTION 1 LCD Display 2 Battery Cover 3 Measure Key 4 Mode Key 5 Sensor 6 Stand...

Page 5: ...tions carefully for best results 1 Press the Meas key to power on the thermometer 2 Always keep the sensor lens and the light cone clean before measurement Please refer to the Clean and Storage sectio...

Page 6: ...er attached on the rear side of the thermometer Warning This Infrared Thermometer TN1 is calibrated for Emissivity 0 95 only Most of Non metal materials have emissivity near to 0 95 But the infrared e...

Page 7: ...ct It also has a LOCK mode for long term operation of the thermometer Using Minimum Maximum Lock Function Power on the thermometer Use a pin stick to press the Mode key several times until the MIN sta...

Page 8: ...me Press the Mode key several times then use Meas key to confirm the setting The measuring will keep going as long as 60 minutes until the Meas key is pressed There is a metal stand allocated on the r...

Page 9: ...ry for at least 30 minutes before using Changing the Battery This device is supplied with one lithium cell CR2032 x 1 A low battery power symbol will appear on the LCD display when battery needs to be...

Page 10: ...nstalled with the incorrect direction Try to switch off then on one more time if abnormal situation occurred If the situation is still contact your local service center for support Battery may be dama...

Page 11: ...ow the thermometer to rest in an environment for more than 30min at recommended operation temperature The ambient temperature is not within the range between 10 C to 50 C 14 to 122 Allow the thermomet...

Page 12: ...et temperature is too high Please take a new temperature measurement Temperature taken is lower than 22 C 7 6 F Make sure if the target temperature is too low Please take a new temperature measurement...