System description

Introduction

The PDC is designed to function as a stand-alone battery powered instrument. Its
application can be extended by connection to a PC computer running the optional
software package (see separate software user manual). A support stand is
provided to elevate the PDC above structures that might add scattered radiation
to a primary beam measurement.
The PDC has been designed for use with a patient equivalent phantom when
measuring entrance or image receptor doses. This is not included in the system
package.

Construction

The PDC consists of an ion chamber assembly, electronic printed circuit board
incorporating displays, a switch panel and battery.
The components are housed within a protective housing that incorporates a beam
alignment target.

Note

. There are no user accessible components within the housing.

The Ion chamber

The ion chamber is suspended within the protective housing to minimize
disturbance during positioning the PDC in the radiation beam. The chamber
consists of central and outer active areas. Dose measurement uses the central
area only whilst both areas are used for DAP measurement.

The radiation target

A target assembly is provided to assist with alignment to the radiation beam.

Attention

. When making dose measurements the radiation field size must exceed

the red 10cm diameter area.
Where a light field is not available a matrix of radio-opaque markers are provided
for alignment with a fluoroscopic image. The markers are placed at points of
major intercept on the target and are illustrated by red dots.

The Electronic system

The PDC uses a micro controller to manage and process instructions and
measurement data. A common high voltage generator is used to energize both
sections of the ion chamber. The system incorporates three independent
electrometers that are permanently coupled to collector plates of the ion chamber.
Their functions are:
1. Central area
2. Outer area, low sensitivity
3. Outer area, high sensitivity
The electrometer outputs are continuously sampled by the processor.

8

Summary of Contents for PDC

Page 1: ...Model PDC Patient Dose Calibrator Instruction Manual...

Page 2: ......

Page 3: ...E FOLLOWING PATENT PAT NO 7 413 345 Radcal Corporation 426 West Duarte Road Monrovia CA 91016 4591 USA USA 626 357 7921 Fax USA 626 357 8863 email Service radcal com www radcal com Radcal Part MNL PDC...

Page 4: ...n the battery charger is connected Always disconnect the battery charger before using the system Important Throughout this manual the following words are given a specific meaning and the reader should...

Page 5: ...8 Computer connection 9 Preparations for operation 10 Preparation 10 Charging the battery 10 Power on test 10 Stabilization time 10 Operation 11 Switching on the device 11 RESET function 12 TEST funct...

Page 6: ...ile HF communication systems may affect electrical measuring devices The Patient Dose Calibrator was designed to power off if exposed to electro static discharge ESD Protection against electric shock...

Page 7: ...e risk of such insulation failures can be reduced by drying the ionization chamber at an elevated temperature max 50 C A relative air humidity exceeding 80 can cause precipitation of condensate on the...

Page 8: ...ze disturbance during positioning the PDC in the radiation beam The chamber consists of central and outer active areas Dose measurement uses the central area only whilst both areas are used for DAP me...

Page 9: ...conds after the end of a radiation exposure The system performs an automatic test routine when switched on This involves a display test electrometer test battery check and electrometer auto zero The b...

Page 10: ...e PDC battery is completely flat when the charger is plugged in its yellow charging indicator lamp fails to glow until the load on the power supply is reduced after about 30 minutes of charging Power...

Page 11: ...s and a digit test of the display is then performed Thereafter the power on test sequence starts TEST The required time for the power on test is displayed Three test values T1 T3 are then displayed a...

Page 12: ...rror message is displayed At the end of the power on test a battery check is carried out and the percentage state of charge is displayed When the PDC is switched on the TEST function is started automa...

Page 13: ...urements Depress the Mode button until the mGy indicator is illuminated Make sure that the x ray field size is larger than the red 10 cm diameter circle area indication on the PDC cover Set the exposu...

Page 14: ...est values and contact your Radcal representative OVERRATE maximum dose rate is exceeded Change the measuring range to high rate RESET repeat measurement LOW RATE Below minimum dose rate Change the me...

Page 15: ...100 mm2 at center of chamber assembly Chamber polarization potential 300 V Distance between electrodes 10 5 mm Stabilization time 3 min Measuring system Digital resolution Dose area product 0 01 GyAm...

Page 16: ...ply Internal rechargeable battery pack Li ION 2 cells Panasonic CGR18650CF UL 1642 Nominal voltage 7 2 V Nominal capacity 2 25 Ah Maximum charge voltage 8 2 V Maximum charge current 2 5 A Number of ch...

Page 17: ...Support Elevation Stand Foam Elevation Support Stand fits inside the interior of the Carry Case lid 17...

Page 18: ...replacement parts nor shall Radcal be liable for any defective material replaced without Radcal s consent during the period of this warranty Radcal reserves the right to perform warranty services at...

Page 19: ...ber s PDC 01 Product Accessories All Conforms to the following Product Specification EMC EN 61326 1 2013 Group 1 Class B EN 61000 4 2 EN 61000 4 3 EN 61000 4 4 EN 61000 4 6 EN 61000 4 5 EN 61000 4 11...

Reviews: