Model DAPCKP
DAPcheck Plus 
DAP Analyzer

Instruction Manual

Summary of Contents for DAPcheck Plus

Page 1: ...Model DAPCKP DAPcheck Plus DAP Analyzer Instruction Manual...

Page 2: ......

Page 3: ...nder the Following Patent Pat No 7 413 345 Radcal Corporation 426 West Duarte Road Monrovia CA 91016 4591 USA USA 626 357 7921 Fax USA 626 357 8863 email Service radcal com www radcal com Radcal Part...

Page 4: ...c meaning and the reader should be aware of their significance Attention Used where there is a risk of damaging the DAPcheck Plus by incorrect handling or operation error Note Used where information i...

Page 5: ...ing 6 Cleaning 6 Storage conditions of application 6 Disposal 7 System description 8 Introduction 8 Construction 8 The Ion chamber 8 The radiation target 8 Preparations for operation 9 Preparation 9 P...

Page 6: ...e 0 C 50 C Relative air humidity 10 80 max 20 g m3 non condensing Cleaning When cleaning the ionization chamber no abrasive cleaners and no organic solvents or cleaning agents containing solvents e g...

Page 7: ...Disposal The DAPcheck Plus must be disposed of in accordance with the applicable national regulations or returned to the manufacturer A WEEE manual is available from Radcal 7...

Page 8: ...incorporates a beam alignment target Note There are no user accessible components within the housing The Ion chamber The ion chamber is suspended within the protective housing to minimize disturbance...

Page 9: ...measuring system includes an automatic power on test during which the function of the ionization chamber and all electronic systems are checked Stabilization time The ionization chamber is a highly s...

Page 10: ...ed on the desktop automatically AccuGoldExcel is an Excel interface for Accu Gold and the software is installed as an Excel add on A quick start manual can be found under Manuals in the same folder Us...

Page 11: ...10 mGy min 15 x 15 cm field 80 kVp 2 5 mm Al filtration Digital Resolution Dose area product rate 1 Gy m2 min to 0 91 Gy m2 min Rated range of use Tube voltage 40 150 kV Automatic Temperature and Pres...

Page 12: ...Connection to Accu Gold Digitizer One meter disconnect cable with 2 connectors Protection Class acc IEC 60529 IP 41 Weight 2 3 kg 5 1 lbs Dimension L x W x H 350 mm x 410 mm x 35 mm 12...

Page 13: ...Support Elevation Stand Foam Elevation Support Stand fits inside the interior of the Carry Case lid 13...

Page 14: ...14...

Page 15: ...ement parts nor shall Radcal be liable for any defective material replaced without Radcal s consent during the period of this warranty Radcal reserves the right to perform warranty services at its own...

Page 16: ...sole responsibility that the model DAPcheck Plus Measurement System conforms to the following product specifications EMC EN50081 2 EN55011 Class A EN61000 4 2 EN61000 4 3 EN61000 4 8 IEC60601 1 IEC60...