background image

Accu-Gold Touch 

User Guide

16

USB Mode - Using Accu-Gold 2 software
Connect the control unit to a PC using the supplied micro USB cable (or use a standard USB-A to USB-B micro cable) 

and attach any sensor that you would like to measure with.  The maximum USB cable length is 3 meters.  You can use 

a 5 meter cable but you may experience interference issues in some cases.  Over 5 meters requires using an active 

extension (contact Radcal).
Launch Accu-Gold 2 on you computer and select "Connect via USB" on the Start screen.
Refer to the Accu-Gold 2 User Manual when in USB or WiFi Modes.
When the software displays the "Ready" status in the lower left corner of the window you can begin measuring. 

Summary of Contents for Accu-Gold Touch

Page 1: ...ACCU GOLD TOUCH USER GUIDE For use with all Accu Gold Touch Models...

Page 2: ...d to meet future needs The Accu Gold Touch provides the ability to save all of your measurements The Accu Gold Touch system consists of a built in digitizer module whose function is to transform the a...

Page 3: ...ed at two different viewing angles 35 60 Touch Screen Multisensor Auxiliary and mA Connectors ON OFF Ion Chamber and Dose Diode Connectors USB Flash Drive Connection Measurement Export and Firmware Up...

Page 4: ...ch can be used like a traditional handheld meter with an easy to use interface The Pro version will have a starting menu to be able to also use the Touch in USB and WiFi modes On the start screen clic...

Page 5: ...easurements Delete current measurement Four outputs can be viewed on the display at one time Sensor legend Color Coded MS Multisensor IC Ion Chamber DAP DAP Chamber DAPChk DAPcheck Plus DD Dose Diode...

Page 6: ...Accu Gold Touch User Guide 6 Waveforms Start trigger light grey kV MS or Dose dashed Dose Rate solid End of pulse trigger point...

Page 7: ...ct High if Std causes false triggering Dental Mode Designed to accommodate the initial pre heat kV pulses commonly found with some dental generators The dose remains integrated over the entire exposur...

Page 8: ...sor you use A diagnostic multisensor will only show the diagnostic calibrations an ion chamber will show none Some calibrations require that the multisensor have a particular built in calibration and...

Page 9: ...Accu Gold Touch User Guide 9 Anode Filter Calibrations Available continued...

Page 10: ...Gold 2 software V2 46 or later Copy the exported file to a PC with AG2 on it Double click this file and it will open in the AG2 software The Touch only shows the last 300 measurements but will export...

Page 11: ...ple sensors can be combined on a single page showing the results of interest without having to change pages Touch any measurement result A page opens up that lets you replace the result shown with ano...

Page 12: ...S2 Average peak voltage Ion Chamber Dose IC Dose Integral Dose Pulse Dose Pulse Dose Rate IC Dose Rate Dose Rate Max Maximum dose rate encountered Temperature Temperature measured by ion chamber The a...

Page 13: ...m of mA values during the region of interest is the value of mAs Current mA Average mA is mAs divided by the width of the region of interest Charge Pulse mA Light Sensor Illuminance LS Measured illumi...

Page 14: ...easurement when the exposure is complete 9 At any time you can review outputs of the current measurement or previous measurements Make additional exposures as necessary Note All measurements are autom...

Page 15: ...solid when complete The Touch can be used while being charged and the charger may be left connected indefinitely If the battery has been exhausted there will be a small delay while it charges up enoug...

Page 16: ...e maximum USB cable length is 3 meters You can use a 5 meter cable but you may experience interference issues in some cases Over 5 meters requires using an active extension contact Radcal Launch Accu...

Page 17: ...screen When the software displays Ready you can begin to measure If the connection fails go to the windows connection manager and select TouchXX XXXX and click connect If a PIN is requested click the...

Page 18: ...single cell Battery Life 6 Hours Charging Time 5 5 Hours maximum to fully recharge Charger Radcal part number PRS PSA10F 050 5V 2 0A Input 90 to 264 VAC 47 to 63 Hz AC power supply blades internationa...

Page 19: ...furnishing replacement parts nor shall Radcal be liable for any defective material replaced without Radcal s consent during the period of this warranty Radcal reserves the right to perform warranty s...

Page 20: ...t Duarte Road Monrovia CA 91016 4591 USA USA 626 357 7921 Fax USA 626 357 8863 email sales radcal com www radcal com Radcal Part MNL AGTQS 4094382 REV G Firmware V1 45 on Printed June 2020 Copyright 2...