1260-100X User Manual
Racal Instruments
©
2005
Troubleshooting C-1
Appendix C
TROUBLESHOOTING
INTRODUCTION
If you have difficulty in using the 1260-100X Adapt-a-Switch
Carrier or its plug-ins, refer to the following troubleshooting table.
If you are still unable to resolve the problem, contact one of the
Racal Customer Support offices listed in
Chapter 5: Support
Offices
.
TROUBLESHOOTING TABLE
PROBLEM
POSSIBLE CAUSES
WHAT TO DO
No response from plug-in via
VXIbus.
•
Option-01T not installed in
the carrier controlling the
1260-100X, or installed in
incorrect module.
•
Module
addresses
conflicting.
•
Plug-in not fully seated.
•
Communication problem with
Option-01T, 1260-100X
Carrier, or plug-in.
•
Option-01T must be installed in
left-most carrier or 1260-Series
module in group. See
Option-01T
in
Chapter 2:
Getting Started
.
•
Please note that currently 1260-
100X does not have an Option-
01T feature. It requires another
carrier with an Option-01T to
control it.
•
Ensure that each plug-in and
1260-Series module has its own
unique module address. See
Module Address Switch
in
Chapter 2: Getting Started
.
•
Ensure that plug-in is firmly
engaged.
•
See special troubleshooting
procedure below, entitled
“Communications
Troubleshooting”