G
B
MAINTENANCE
The x:panda cushion is removable and the cover can be washed at max. 40°. The condition of
the foam cushion should be checked at least every six months and replaced if necessary. It is
important to keep the frame clean to maintain the validity of the chair.
The carer should check and oil the swing away parts occationally - at least every 6
months.
Use a dry cloth, on a weekly basis, to wipe off filth and dirt from the wheels. Do not use any
cleaning materials containing chlorine or methylated spirit.
The x:panda has to be checked and tightened by a qualified technician at least every 6
months to avoid failures.
Do not expose the gasspring to
pressure, high temperatures or perforation.
Any adjustments and addition of accessories which are not described in this manual, must
only be made by authorized R82 personnel. Vests and belts must be mounted according to
the instruction provided with the products.
The carer should check and refill the tyres for air at least once a month. Air pressure:
High-low:x outdoor frame
.
Front: 35 PSI/250 kPa, rear: 40 PSI/2,5 bar
Wheelchair frame
.
Wheels: 110 PSI/250 kPa
Pushchair frame
.
Front: 36 PSI/2,5 bar, rear: 36 PSI/2,5 bar