R-Can Sterilight S12Q Installation Instructions And Owner'S Manual Download Page 3

•  If treating the entire house, install the disinfection system before any branch lines. Ideally, your disinfection system should

be the last treatment your water receives prior to use.

•  A 5 micron sediment filter must precede the disinfection system.

1.

Remove the disinfection system from the shipping carton. For shipping purposes, the UV lamp is packed in a separate

tube. Set the lamp aside for use later. The disinfection system should be mounted in the horizontal position, with the
inlet/outlet ports facing up. If the system must be installed in the vertical position, make sure the inlet port is the one at the
bottom of the system. Mount the unit in a clear space with at least 36” (91.5 cm) of space at the lamp end to facilitate
lamp and or quartz sleeve removal. Fasten the disinfection system to a suitable mounting platform with secure fasteners.

2.

It is recommended to install suitable flow regulatiors in order that the flow rate not exceed the manufacturers 

recommended flow rating. The use of by-pass and shut-off valves are recommended for emergency use of untreated
water when your disinfection system is being serviced.

Please Note: 

When the UV unit is by-passed for service, the com-

plete water system must be sterilized once again with chlorine to destroy any contamination that may have passed during
by-pass. Apply two turns of Teflon tape around the port threads to ensure a tight joint.

3.

Very carefully

slide the UV lamp into the UV chamber making sure the lamp pins are accessible for connection with the

lamp connector cable. Attach the lamp connector to the lamp pins and press fit into aluminum gland nut.

4.

When all plumbing connections are made,

slowly 

turn on the water supply and check for leaks. The most likely cause for

leaks is from the o-ring seal. In case of a leak, shut water off, drain cell, remove the retainer nut, wipe the o-ring and
threads clean and re-install.

5.

Once it is determined that there are no leaks, plug the disinfection system into the ground fault interrupter, and check to

see if the UV lamp is illuminated.

NEVER LOOK DIRECTLY AT THE BURNING UV LAMP

. Secure the lid, and allow the water to

run for a few minutes to clear any air or dust that may be in the cell.

PLEASE NOTE:

When there is no flow, the water in the cell will become warm, as the UV disinfection system lamp is always

on. To remedy this, run a cold water tap anywhere in the house for a minute to flush out the warm water.

•  Always disconnect power before performing any work on the ultraviolet disinfection system.
•  Regularly inspect your disinfection system unit to ensure that the UV light is still glowing and that its LED indicator is still   

glowing green.

•  Replace the UV lamp annually ( or biennially if seasonal cottage use) to ensure a high bacteria and virus kill 

rate.

•  Always drain the UV cell when closing a cottage or leaving the unit in an area subject to freezing 

temperatures.

A. UV LAMP REPLACEMENT

1

. There is 

NO

need to disconnect the disinfection system from the water supply or to drain the water from the reactor 

chamber, nor do you need to remove the reactor chamber from the disinfection system. Lamp replacement is a quick and
simple procedure requiring only a slotted screwdriver to perform the job. After 9000 hours of continuous 
operation (approximately one year), the UV lamp must be replaced.

2

. Disconnect main power source. Remove lamp connector from gland nut and carefully slide UV lamp slightly out of 

chamber. Remove lamp connector from lamp and fully remove lamp (be extremely careful when handling UV lamp as
they are extremely fragile).

3.

To install the new lamp, carefully remove from protective tube being careful not to touch the lamp “glass” with your

hands. Slide UV lamp into the reactor chamber. Affix lamp connector to the UV lamp, slide  lamp fully into cell, press fit
lamp connector into aluminum gland nut.

B. QUARTZ SLEEVE REPLACEMENT AND/OR CLEANING

1.

Mineral deposits and sediment may accumulate on the quartz sleeve decreasing UV output. Good maintenance of 

filtration equipment will reduce the accumulation of residues. If necessary, remove the quartz sleeve after a few months
and clean with a commercially available scale remover (CLR, Lime-Away, etc.) and lint free cloth. Repeat the process as
often as necessary to keep the quartz sleeve clean.

2.

Disconnect Power to remove the quartz sleeve. Disconnect the UV lamp and remove from the UV chamber. The sleeve

can be removed in the same manner as the UV lamp (through the end of the disinfection system) or by removing the
entire cell. Remove the aluminum gland nuts as well as the two o-ring seals. Carefully slide the quartz sleeve out of the
chamber (either side) being careful not to let the end drop inside the chamber accidentally cracking the quartz. When re-
assembling the quartz sleeve, make sure the sleeve protrudes an equal distance at each end of the cell. Wet the o-rings

22

Operating & Maintenance Instructions:

Summary of Contents for Sterilight S12Q

Page 1: ...TION INSTRUCTIONS AND OWNERS MANUAL Revisied January 2003 P N 520022 425 Clair Road West P O Box 1719 Guelph ON N1H 7X4 Canada t 519 763 1032 f 519 763 5069 e water r can com i www r can com EPA 57987...

Page 2: ...atts than the disinfection system rating may overheat Care should be taken to arrange the cord so that it will not be tripped over or pulled 10 SAVE THESE INSTRUCTIONS improved transient protection im...

Page 3: ...old water tap anywhere in the house for a minute to flush out the warm water Always disconnect power before performing any work on the ultraviolet disinfection system Regularly inspect your disinfecti...

Page 4: ...ndicates that the UV intensity is satisfactory and the unit is in good working order YELLOW indicates that the UV intensity is reduced which could be due to any of the following factors The lamp is lo...

Page 5: ...s well as acting as an interface for an optional remote audible visual alarm package available from your dealer This IR ALARM comes with the IR interface and 15 24 meters 50 of cable for remote monito...

Page 6: ...ge with appropriate five micron cartridge Note check source water supply as fluctuations may occur in source pressure HIGH BACTERIA COUNTS quartz sleeve is stained or dirty clean sleeve with scale cle...

Page 7: ...SY YM MP PT TO OM MS S R RE EM MA AK KS S P PO OW WE ER R O ON N R RE ED D L LE ED D L LA AM MP P S ST TA AT TU US S G GR RE EE EN N L LE ED D A AU UD DI IB BL LE E A AL LA AR RM M U UV V L LA AM MP...

Page 8: ...ens or replace if required Replace lamp with manufacture replacement Replace sensor with manufacturer replacement Always Wet Sensor Lens prior to inserting sensor RED ON ON UV Sensor has not detected...

Page 9: ...40UV 2 CIRCUIT BOARD for 230V unit RN 001 ALUMINUM RETAINING NUT CP 1240 CIRCUIT BOARD COVER CS 001 Q CORD SET 115V CS EU Q EUROPEAN SCHUKO CORD SET 230V DP750 RETAINING NUT PLUG SP 3 8 3 8 DRAIN PLUG...

Page 10: ...XI IM MU UM M O OP PE ER RA AT TI IN NG G P PR RE ES SS SU UR RE E 8 62 bar 125 psi 8 62 bar 125 psi 8 62 bar 125 psi 8 62 bar 125 psi 8 62 bar 125 psi A AM MB BI IE EN NT T T TE EM MP PE ER RA AT TU...

Page 11: ...TION ET MANUEL D UTILISATION r vis janvier 2002 P N 520022 EPA 57987 CN 001 pour la s curit de votre eau 425 Ouest Clair Road C P 1719 Guelph ON N1H 7X4 Canada t 519 763 1032 f 519 763 5069 e water r...

Page 12: ...filtrage IEM am lior circuit d alarme am lior pour signalisation d une d faillance de fusible allumage am lior de la lampe protection contre les baisses de tension transitoires carte de circuits avec...

Page 13: ...ue le t moin vert DEL reste illumin Remplacer la lampe UV chaque ann e ou intervalles de deux ans pour une utilisation saisonni re pour garantir un taux lev de destruction des bact ries et virus Vidan...

Page 14: ...lectrique r duisent la puissance de sortie de la lampe La lampe retournera la normale lorsque l lectricit retournera sa tension totale Remarque le syst me de contr le ne fonctionnera pas pendant des p...

Page 15: ...REMENT UNE ANALYSE BACT RIOLOGIQUE DE L EAU PAR UN LABORATOIRE DE TEST RECONNU Le fabricant garantit que les syst mes lectriques et les composants du syst me de d sinfectionUV sont exempts de vice de...

Page 16: ...p ratif que l eau effluente re oive un traitement de choc au chlore apr s que l eau quitte le syst me de d sinfection le syst me de d sinfection doit avoir un syst me de distribution sans bact ries po...

Page 17: ...SU UE EL LS S R RE EM MA AR RQ QU UE ES S T TE EN NS SI IO ON N D DE EL L R RO OU UG GE E I IN NT TE EN NS SI IT T U UV V D DE EL L V VE ER RT TE E A AL LA AR RM ME E S SO ON NO OR RE E L LA AM MP PE...

Page 18: ...t Remplacez le d tecteur avec un d tecteur de rechange du fabricant Mouillez toujours la lentille du d tecteur avant de l ins rer ROUGE ALLUM E ALLUM E Le d tecteur UV n a pas d tect une intensit UV s...

Page 19: ...ARTE DE CIRCUIT FOR 230V UNIT RN 001 CROU D IMMOBILISATION CP 1240 COUVERCLE DE LA CARTE DE CIRCUIT CS 001 Q CORDON D ALIMENTATION 120V CS EU Q CORDON D ALIMENTATION EUROP EN 230V DP750 BOUCHON SP 3 8...

Page 20: ...N 8 62 bar 125 lb po2 8 62 bar 125 lb po2 8 62 bar 125 lb po2 8 62 bar 125 lb po2 8 62 bar 125 lb po2 T TE EM MP P R RA AT TU UR RE E A AM MB BI IA AN NT TE E 2 4 C 36 104 F 2 4 C 36 104 F 2 4 C 36 10...

Reviews: