2014-09-30
Page 12 of 34
R-Biopharm AG
Getting started
The proper placement of the RIDA
®
CUBE SCAN Laboratory Photometer
Place the Instrument on a dry, clean, stable and level surface.
Make sure the Instrument has at
least 10 cm of table surface and clearance on each side and that the Instrument can be easily
disconnected from the power source. Allow the Instrument to acclimate to the ambient room
temperature
before operating it.
The Instrument can be damaged by:
•
Condensing humidity and water
•
Heat and large temperature fluctuations
•
Direct sunlight
•
Vibrations (e.g. from centrifuges and dishwashers)
•
Electromagnetic radiation
•
Electrostatic discharge
Transport lock
Upon first use the white transport lock is to be removed from the door of the Instrument.
Note:
keep both the transport lock and the original packaging. In case of a defect the
Instrument is to be returned in the original packaging (see p. 7) with the transport lock
installed to prevent any damage during transportation!
Connecting the power supply
.
RIDA
®
CUBE SCAN Instrument Interfaces
Power supply
USB Socket
for connecting a
PC (for data transfer)
Tablet PC-Lock eyelet
Bluetooth on/off
switch
(left: Bluetooth OFF)
USB Port
for connecting a
Tablet PC
RS 232
socket for connecting a
printer, a barcode reader or a
PC (for data transfer)
-
Connect the power cable to the power supply unit.
-
Insert the plug from the power supply unit into the power socket on the back of the Instrument
-
Plug the power cable into the wall socket.
Always connect to the proper supply voltage. The power supply voltage must comply
with the regulations cited in the technical specifications on page 31. Ensure a properly
installed electrical grounding. The Instrument is to be operated only using the power
supply unit provided.