4-16 Repair/Replacement and Calibration
Cardiac Stress Treadmills Service Manual
5. Remove the two hex-head bolts holding the wheels to the rack gears.
Note the arrangement of washers used as spacers.
6. Remove the grade pot and bracket assembly (see step 6, “Replacing
the Grade Potentiometer (Pot) on page 4-14).
7. Rotate the grade pot until a positive number appears in the grade
display window of the stress monitor system. (Step 9 will not work
until the number is positive.)
8. (Requires two people) Decrease grade until rack gears start bouncing
on the pinion shaft. Hold the limit switches closed to let the rack gear
travel beyond its normal range.
9. Turn off and unplug the treadmill.
10. Lift rack gears straight out top.
Reassembling the Rack Gears
Install new rack gears at the same time so that they extend an equal
distance through the headframe.
1. Restore power to the treadmill.
2. Loosen the setscrew that secures the grade potentiometer gear and
remove the gear.
3. Insert rack gears into headframe.
4. Decrease the grade until the gears bounce two or three times, then
increase the grade. This should cause both rack gears to mesh in
exactly the same place.
(step 11)
Rack Gears
3/8-16 x 2.5OL
(step 5)
Summary of Contents for TM55
Page 1: ...Cardiac Stress Treadmills Service Manual P N 042099 001 Rev A...
Page 4: ......
Page 8: ......
Page 14: ......
Page 23: ...Cardiac Stress Treadmills Service Manual Troubleshooting 3 9 Grade Problems Flow Chart...
Page 24: ......
Page 52: ......
Page 56: ......
Page 71: ...Cardiac Stress Treadmills Service Manual PartNumbers C 7 27 26 27 13 18 15 26...
Page 72: ...C 8 Part Numbers Cardiac Stress Treadmills Service Manual 30 29 31 28...
Page 74: ......
Page 78: ......
Page 80: ...E 4 Drawings Cardiac Stress Treadmills Service Manual Wiring Diagram High Voltage Treadmill...