CONTROLLER ALARM

Controller Alarm-  

If the unit exceeds the high limit alarm value, the controller will 

stop the heating process and display the ALM indicator on the controller until the unit is 

manually reset. 

To reset alarm condition

- Power the unit OFF and allow it to cool down to room 

temperature before powering ON. If the problem persist contact technical support.    

465

ALM

To execute an auto-tuning you will need to:
     Select the desired set-point temperature to tune the unit. 

Press and HOLD 

to see 

PASS

AUTO-TUNING THE UNIT

FAST

 Tuning

P

F

P

F

Your unit’s controller has been tuned from our factory to 

achieve set-point temperatures at a faster rate. However, 

you can auto-tune your unit’s controller to work best with 

your application and set-points. The two auto-tuning options 

available are:

Press      to input 

passcode

1

2

FULL

 Tuning

This option prioritizes speed over accuracy, allowing you to 

reach set-point temperatures faster. 

This tuning option gives priority to accuracy over speed. 

PASS

Once the desired set-point temperature has been set, 

you will need to input the factory set passcode to access 

the tuning parameter.  

SP

CONTROLLER LOCK

Your unit’s controller allows you the option to “lock” the controller buttons to prevent any changes to the parameter values. 

You can do this by following the sequences shown below: 

1. Press         (3) �mes to see 

PASS

2. Enter the password of

1111

  by holding

3. Press and hold         un�l the display 

CALIB

 

4. Press         (3) �mes to see

 

PROT

5. Change the 

4

 

to 

5

 

by pressing

 

6. Press         again (4) �mes to return to temperature

ENTER PASSWORD

CALib Prompt

Prot PROMPT

PASSWORD PROMPT

CHANGE 4 TO 5

RETURN TO 

TEMPERATURE

PASS

1111

CALIB

PROT

5

125

During the tuning process, it is common to see the 

temperature fluctuate by a few degrees above and 

below the setpoint.
Allow the unit some time to reach temperature stability 

after the tuning process has been completed.

IMPORTANT:

3

ATUN

Select either the 

FAST

 or 

FULL

 tuning option to 

be executed. Once a selection has been made exit 

tuning menu. The indicator 

AT

 

will light up and remain 

ON during the tuning process.

P

F

Press     ONCE to 

see 

ATUNE

.

Continue to 

3

Press      or       to

Increase / Decrease

P

F

P

F

Press      3 times to 

return to chamber 

temperature 

P

F

F

Press       once to 

see 

SP

PPP

P

F

F

F

F

Factory set passcode 1111

P

P

P

F

Press      to select 

tuning option

P

F

F

F

     Although there are other tuning options available, 

we recommend only using the two options mentioned 

above. Any other tuning option performed will affect 

the unit’s performance and accuracy.  

Press and HOLD to 

or      to exit tuning 

menu

P

PAGE 4