NOVUS N1020

AT

OPERATION 

INDICATORS 

CONTROLLER

KEYS 

PROCESS VALUE

RUN  

ON - Control is enable to run. 

(This light remains ON during use)

OUT

ON- Indicates normal activation of outputs. 

FLASHING - Indicates normal operation

AT Indicator 

ON- Indicates that auto-tuning is in process. 

(See Auto-Tuning the oven for more information).  

CONTROLLER KEY DESCRIPTION

P

F

Menu/Advance - 

Use to advance to successive parameter or menu  

Increase 

- Use key to increase the parameters value.

Decrease 

- Use key to decrease the parameters value.

Special Function 

- Use key to return and exit. 

OPERATION INDICATORS

 

300

OUT

RUN

ALM

ALM  

ON - Indicates an alarm condition when the process 

value is above the defined alarm value. 

SET-POINT CHANGE  

TEMPERATURE UNITS 

To change the temperature set-point of your oven, use the 

indicated arrow keys to increase or decrease temperature.

To change the oven’s operational unit to read either 

degrees Celsius or degrees Fahrenheit you will need to:

UNIT

SP

TEMPERATURE OFFSET / CALIBRATION

Your unit has been calibrated at our factory using a 

NIST-certified temperature instrument. 
Over time a temperature adjustment (offset) may be neces-

sary to maintain the unit’s controller temperature reading 

consistent with the interior temperature. To maintain this 

temperature accuracy, we recommend verifying twice a year 

that there is a temperature consistency between the control-

ler and the chamber using a known accurate temperature 

measuring device.  

     Place a trusted, preferably certified digital temperature  

     probe at the center of the unit’s chamber.
     Record the temperature reading at the controller and at 

     the center of your unit’s chamber.
     Access the 

OFFS 

parameter and input the tempera- 

     ture difference from your controller to your digital probe.   

11

2

3

To navigate the controller's menu, please refer to the easy-to-use Menu Guide below:  Menu setting changes can be done 

quickly with our 4-button digital microprocessor. Your incubator's controller allows you to perform the following: 

Set the operating set-point temperature
Select between degrees Fahrenheit or Centigrade
Calibrate your unit to your independent temperature sensing device. 

Auto-tune your Incubator for maximum efficiency
Lock the set-temperature

M E N U   G U I D E

P

F

P

F

Press      ONCE to 

see 

OFFS

P

F

Press      3 times to 

see 

PASS

Press      to input 

factory passcode 

P

F

Press      or      to input 

temperature difference

1111

P

F

Press       ONCE to

exit calibration 

Once a temperature adjustment has been made, allow 

the unit some time to stabilize before making any 

subsequent adjustements.

The temperature difference between the controller and 

your independent probe can be a positive or negative 

number.

To execute an OFFSET you will need to:

Factory set passcode 1111

To change the offset value on your incubator: 

Press      or       to

Increase / Decrease

P

F

P

F

Press      3 times to 

return to chamber 

temperature 

P

F

F

Press       once to 

see 

SP

PPP

P

Press      or       to

select unit

P

F

P

F

Press      2 times to 

return to chamber 

temperature 

P

F

Press       2 times to 

see 

UNIT

PPP

P

P

F

F

F

F

F

F

F

F

F

P

P

PAGE 3

   

Menu Functions Guide