NOVUS N1020
AT
OPERATION
INDICATORS
CONTROLLER
KEYS
PROCESS VALUE
RUN
ON - Control is enable to run.
(This light remains ON during use)
OUT
ON- Indicates normal activation of outputs.
FLASHING - Indicates normal operation
AT Indicator
ON- Indicates that auto-tuning is in process.
(See Auto-Tuning the oven for more information).
CONTROLLER KEY DESCRIPTION
P
F
Menu/Advance -
Use to advance to successive parameter or menu
Increase
- Use key to increase the parameters value.
Decrease
- Use key to decrease the parameters value.
Special Function
- Use key to return and exit.
OPERATION INDICATORS
300
OUT
RUN
ALM
ALM
ON - Indicates an alarm condition when the process
value is above the defined alarm value.
SET-POINT CHANGE
TEMPERATURE UNITS
To change the temperature set-point of your oven, use the
indicated arrow keys to increase or decrease temperature.
To change the oven’s operational unit to read either
degrees Celsius or degrees Fahrenheit you will need to:
UNIT
SP
TEMPERATURE OFFSET / CALIBRATION
Your unit has been calibrated at our factory using a
NIST-certified temperature instrument.
Over time a temperature adjustment (offset) may be neces-
sary to maintain the unit’s controller temperature reading
consistent with the interior temperature. To maintain this
temperature accuracy, we recommend verifying twice a year
that there is a temperature consistency between the control-
ler and the chamber using a known accurate temperature
measuring device.
Place a trusted, preferably certified digital temperature
probe at the center of the unit’s chamber.
Record the temperature reading at the controller and at
the center of your unit’s chamber.
Access the
OFFS
parameter and input the tempera-
ture difference from your controller to your digital probe.
11
2
3
To navigate the controller's menu, please refer to the easy-to-use Menu Guide below: Menu setting changes can be done
quickly with our 4-button digital microprocessor. Your incubator's controller allows you to perform the following:
Set the operating set-point temperature
Select between degrees Fahrenheit or Centigrade
Calibrate your unit to your independent temperature sensing device.
Auto-tune your Incubator for maximum efficiency
Lock the set-temperature
M E N U G U I D E
P
F
P
F
Press ONCE to
see
OFFS
P
F
Press 3 times to
see
PASS
Press to input
factory passcode
P
F
Press or to input
temperature difference
1111
P
F
Press ONCE to
exit calibration
Once a temperature adjustment has been made, allow
the unit some time to stabilize before making any
subsequent adjustements.
The temperature difference between the controller and
your independent probe can be a positive or negative
number.
To execute an OFFSET you will need to:
Factory set passcode 1111
To change the offset value on your incubator:
Press or to
Increase / Decrease
P
F
P
F
Press 3 times to
return to chamber
temperature
P
F
F
Press once to
see
SP
PPP
P
Press or to
select unit
P
F
P
F
Press 2 times to
return to chamber
temperature
P
F
Press 2 times to
see
UNIT
PPP
P
P
F
F
F
F
F
F
F
F
F
P
P
PAGE 3
Menu Functions Guide