Upon start-up all Quincy Lab Bench Ovens will display the current
chamaber temperature (green LEDs) and begin the heating process to
the factory default temperature set-point (orange LEDs) which varies
depending on your oven’s model.
To change the factory default set temperature simply do the following:
Turn the power/recirc. fan switch to the up position.
The blower will begin to circulate chamber air and the temperature controller will power ON.
Turn heater switch to the up position (heat).
Units heat elements will activate and begin heating.
If heater switch is not in the ON position the unit will not heat.
Digital Controller Set-point Temperature.
The controller will begin its normal process to reach the factory default temperature set-point. The heat cycle indicator
light on the controller will activate and cycle on and off with the heaters to reach and maintain the set temperature.
Each shelf will support a distributed load of 100 lbs. maximum.
Do not exceed a combined total of 300 lbs. within the oven at one
time. Avoid placing articles on the oven floor. Instead, use a shelf
at the lowest adjustable position.
Install adjustable shelf by first placing the shelf bracket rivets into the
corresponding keyhole supports located on each inner side of the oven.
Orientate the bracket in the "down" or " L" position. This position guides
the shelf in and out and protects the side wall from being scratched.The
bracket may also be placed in the "up" or " " position if slightly more
interior clearance is needed. Place the shelf on the brackets as shown.
(FIG. 3)
Care should be taken when removing articles from the oven.
Don't pull the shelf out when removing heavy loads.
The shelf
is not secured and loads can tip and fall forward.
Actual Chamber Temperature
Temperature Setpoint
Heat Cycle Indicator
*
All control parameters and other control variables have already been factory set for your model's temperature range and
capacity to provide heating and performance characteristics for a broad range of applications. Your Bench Series ER oven
is ready for operation.
Shelf Installation and Use
DIGITAL MICROPROCESSOR
Temperature Controller
General Operation
NORMAL OPERATION
ADJUSTING CONTROLLER’S SETPOINT TEMPERATURE
*
TO INCREASE TEMPERATURE
SET POINT
*
Press and Hold the and
press the arrow to increase
Set-point temperature
*
*
*
TO DECREASE TEMPERATURE
SET POINT
Press and Hold the and
press the arrow to decrease
Set-point temperature
Star
Key
Down
Arrow Key
UP
Arrow Key
Typically, the oven will need to cycle at the set temperature for a minimum of 20 minutes before it will
achieve equilibrium and becomes stable
(see stability specs. on page 1).
PAGE 3
FIG. 3