Quest Technologies M-27 Instruction Manual Download Page 23

 

between TWA and Dose is that TWA is expressed in decibels and Dose is expressed as a percent.

 

 

TWA is equal to a continuous level which if constant for eight hours would produce the same 

noise dose as the dose accumulated over the run time. The average integrated sound level (Lavg) will be 
equal to TWA at eight hours. If the time is less than eight hours, then TWA will be less than Lavg; TWA 
will be greater than Lavg if time is greater than eight hours. 

 

The M-27 automatically computes the TWA of a worker if the unit is worn by the worker for a full 

workday. Even if the worker does not work exactly eight hours the TWA measurement will read the correct 
TWA for that worker. 

 

Lavg can be used with short term sampling to estimate the worker's TWA. If the noise exposure 

during the sample period is representative of the noise throughout the workday and the workday is eight 
hours, then the TWA for that worker is equal to the Lavg measured with the short term sample. 

 

Estimating TWA from Lavg 

 

A short term sample can be used to estimate worker exposure even if the workday is other than 

eight hours by adding or subtracting a correction factor from the Lavg reading. 

 

As an example, if a worker has a Lavg of 85.50 dB after a ten minute sample, and the sample is 

representative of the noise throughout the workday, and the workday at that position is 7:15 then from 
Table 4 under a 5 dB exchange rate find -.71 dB and subtract it from 85.50 to compute a TWA of 84.79 dB.

 

 

Accuracy of Readings 

 

There is a tendency to overestimate the accuracy of digital readings. In the M-27, some values 

such as Lavg and TWA are computed to 0.1 dB. The absolute accuracy of the reading is not 0.1 dB. (See 
specifications for accuracy and tolerances.) However, the resolution of 0.1 dB is useful in determining the 
sample time required for a short term sample. As an example, if Lavg reading is increasing .3 dB every 
second then additional sample time may be required. If the Lavg reading remains stable within a few 0.1 
dB's then the sample time may be long enough. 

 

Hints and Good Practices 

 

The M-27 is in the pause mode after the battery is replaced, when

 

 

Summary of Contents for M-27

Page 1: ...Microphone Replacement 18 Calibration and Maintenance 18 Daily Calibration 18 Altitude Corrections 19 OPERATING PROCEDURE 19 As a Personal Noise Dosimeter 19 As an Area Monitor 20 As a Survey or Event...

Page 2: ...r is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connec...

Page 3: ...75 80 85 or 90 dB 2 ranges of 35 126 or 50 146 dB A or C frequency weighting Fast or slow time constant Upper limits of 90 115 or 140 dB 12 or 24 hour clocks 8 different unit numbers Text graph or bot...

Page 4: ...IME data If the total time is less than 30 minutes the co lon will blink if the key is held down and the time shown is in minutes and seconds If the colon stays on the time shown is in hours and minut...

Page 5: ...n it should be rechecked PAUSE Resets the calibration level Hold key down until the RESET display shows CAL Using a Printer The printout Figure 2 can include a heading a summary of the data event data...

Page 6: ...hing printers or computers without destroying data See Printer Setup page 32 NOTE Internal switches 1 and 2 are read only when the unit is RESET SPECIFICATIONS Standards ANSI S1 4 1983 type 2 ANSI S1...

Page 7: ...ing the unit off will not destroy the internal memory A water and dust resistant cover protects the unit from the elements and against tampering The unit is small enough to be placed in a shirt pocket...

Page 8: ...e The same as the pause mode except the data necessary to compute Lavg Leq Dose TWA SEL Peak Max Run Time Upper Limit as well as Event data is accumulated Compute Mode If a key has been pressed in the...

Page 9: ...y Test Press any key If the battery is low LOB AT will be displayed in the upper left corner of the display If the battery has at least 8 hours life LOBAT will be off If LOBAT is displayed after a day...

Page 10: ...the microphone Carefully insert the microphone in the microphone adaptor and place the adaptor in the calibrator Make sure the microphone is inserted all the way in the adaptor and the adaptor rests...

Page 11: ...ck of unit and clip microphone to the microphone bracket The microphone clip may be removed and the microphone screwed to the bracket before the bracket is mounted to the case see Figure 7 3 Mount the...

Page 12: ...nstrument the microphone should be mounted on the MB 1 microphone bracket see Figure 7 It may either be clipped to the bracket or the clip removed and the microphone screwed to the bracket The bracket...

Page 13: ...d or only selected data if desired It is possible to change the printout in the middle to obtain for instance a 10 minute histogram for part of the time and a 1 minute histogram for another The M 27 r...

Page 14: ...hange the time base on the histogram in the middle of a printout see Figure 14 for example and page 31 for details If only TEXT was selected on the internal switches the graph portion would be omitted...

Page 15: ...ributions two with a Time constant and two without If the M 27 is used for community or 3 dB industrial monitoring the printout will contain much more information than would a data logger with a singl...

Page 16: ...han 25 feet long and may be more than is needed If only the data for hearing conservation is required press CODE HL3 until the LTL annunciator is displayed Pressing PRINT will reduce the data by one h...

Page 17: ...ally ON graph and text Positions 5 and 6 control the last 2 characters line Most printers work with both ON Positions 7 and 8 are undefined for parallel Computer Connections The M 27 can send its data...

Page 18: ...dB a low threshold level of 80 dB 115 dB upper limit level A frequency weighting the 50 dB to 146 dB range a SLOW time constant and a 12 hour clock If the display is a 3 digit code then the first num...

Page 19: ...Off On 4 dB On Off 5 dB On On 6 dB SWITCH 2 Positions 1 2 3 4 HTL LTL Off Off Off Off 70 dB none Off Off Off On 70 dB none Off Off On Off 75 dB none Off Off On On 80 dB none Off On Off Off 85 dB none...

Page 20: ...code in the first digit column in Table 2 The clock is 12 hours The 0 means the time constant of weighting is A the range is 50 146 the upper limit is 115 dB and the time constant is Slow The first E...

Page 21: ...e changed when the unit is operating the change will be ignored Switch 3 is read at the start of a printout It may be changed to allow printouts to different printers or computers without destroying t...

Page 22: ...TL Dose and TWA If Run Time is less than eight hours TWA is less than Lavg if Run Time is more than eight hours TWA is greater than Lavg The SEL is greater than Leq The 3 dB Dose is greater than the L...

Page 23: ...ple Estimating TWA from Lavg A short term sample can be used to estimate worker exposure even if the workday is other than eight hours by adding or subtracting a correction factor from the Lavg readin...

Page 24: ...rion level Peak noise levels greater than the range of the instrument are clipped and act as if they were at the maximum level of either 146 dB on the 50 dB range or 126 dB on the 30 dB range Minimum...

Page 25: ...ll for a new RA number Send it prepaid and properly packed in the original shipping carton directly to Quest Technologies 1060 Corporate Center Drive Oconomowoc WI 53066 U S A Repair or replacement wo...

Reviews: