APPENDIX I: Electromagnetic Compatibility Declaration
Quell is intended for use in the electromagnetic
environment specified below. The user should
ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions;
CISPR 11
Group 1
Quell uses RF energy only for its internal function. Its
RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions;
CISPR 11
Class B
Quell is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low voltage
power supply network that supplies buildings used for domestic purposes.
Harmonics
Current
Emissions IEC
61000-3-2
Class A
Fluctuations
and Flicker IEC
61000-3-3
Complies
Electrostatic
Discharge
Immunity (ESD);
IEC 61000-4-2
±6kV contact;
±8kV air
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the humidity should be at least 30%
Radiated RF;
IEC 61000-4-3
3 V/m; 80 MHz
to 2.7 GHz
3 V/m compliance level
Electric
Fast
Transient
Burst
Immunity
IEC 61000-4-4
±2kV power leads Mains power quality should be that of a typical domestic establishment.
Fast Surge
Immunity
IEC 61000-4-5
±2kV Common
mode
Radio Frequency
Common Mode
Immunity
IEC 61000-4-6
3 VRMS (80%
AM at 2 Hz;
150kHz to
80 MHz
Power Frequency
Magnetic Field
Immunity
IEC 61000-4-8
3 A/m
50 Hz
Voltage Dip
and Interrupt
Immunity
IEC 61000-4-11
Complies
Mains power quality should be that of a typical domestic establishment.
APPENDIX I
Electromagnetic Compatibility Declaration