Quantum Turret 6 Manual & Product Overview Download Page 15

15

 

 

 
[F1 ID ?] 

What

 

is

 

the

 

ID

 

number

 

of

 

the

 

sample

 

holder

 

being

 

controlled?

  

......[F1 ID 14] 

Sample

 

holder

 

is

 

a

 

t2

 

or

 

other

 

single

 

cuvette

 

sample

 

holder.

 

 

 

 

 

 

 

Assigned

 

Identities:

 

 

ID

 

=

 

00

 

 

specialty

 

sample

 

holder

 

(see

 

command

 

class

 

14)

 

 

14

 ‐ 

t2

 

 

24

 ‐ 

t2x2

 

 

34

 

 

reserved

 

for

 

turret

 

or

 

linear

 

multi

sample

 

holder

 

 

 

 

 

 

 

 

 

 

2.

 

Controller

 

Firmware

 

Version

 

 
[F1 VN ?] 

What

 

is

 

the

 

version

 

number

 

of

 

the

 

controller

 

firmware?

 

......[F1 VN 1.00] 

The

 

controller

 

is

 

operating

 

firmware

 

version

 

number

 

1.00.

 

 

 

 

 

 

 

 

 

 

 

 

 

3.

 

 

Stirrer

 

 

[F1 MS ?] 

What

 

is

 

the

 

maximum

 

stirrer

 

speed?

 

......[F1 MS 2500] 

The

 

maximum

 

stirrer

 

speed

 

is

 

2500

 

rpm.

 

[F1 LS ?] 

What

 

is

 

the

 

lowest

 

stirrer

 

speed?

 

......[F1 MS 300] 

The

 

lowest

 

stirrer

 

speed

 

is

 

300

 

rpm.

 

[F1 SS S 1000] 

Set

 

stirrer

 

speed

 

to

 

1000

 

rpm.

 

[F1 SS S 0] 

Turn

 

stirrer

 

off

 

and

 

set

 

the

 

speed

 

to

 

zero.

 

[F1 SS +] 

Turn

 

stirrer

 

on

 

and

 

set

 

it

 

to

 

the

 

most

 

recent

 

non

zero

 

stirrer

 

speed

 

setting.

 

[F1 SS -] 

Turn

 

stirrer

 

off.

 

[F1 SS ?] 

What

 

is

 

the

 

current

 

stirrer

 

speed

 

setting?

 

......[F1 SS 1000] 

Stirrer

 

speed

 

setting

 

is

 

1000

 

rpm.

 

 

 

 

 

 

 

 

4.

 

 

Temperature

 

Control

 

 
[F1 TC +] 

Turn

 

temperature

 

control

 

on.

 

[F1 TC -] 

Turn

 

temperature

 

control

 

off.

 

 

 

 

 

 

 

 

 

 

 

5.

 

 

Target

 

Temperature

 

 
[F1 TT S 23.10] 

Set

 

target

 

temperature

 

to

 

23.10

 

ºC.

 

[F1 TT ?] 

What

 

is

 

the

 

current

 

target

 

temperature?

 

......[F1 TT 71.32] 

Target

 

temperature

 

is

 

71.32

 

ºC.

 

[F1 TT +] 

Turn

 

on

 

automatic

 

reporting

 

of

 

manual

 

changes

 

to

 

the

 

target

 

temperature.

 

[F1 TT -] 

Turn

 

off

 

automatic

 

reporting

 

of

 

manual

 

changes

 

to

 

the

 

target

 

temperature.

 

Summary of Contents for Turret 6

Page 1: ...19 8606 Tel 509 624 9290 Fax 509 624 9488 E mail quantum qnw com Web www qnw com Turret 6 6 Position Temperature Controlled Cuvette Turret for UV Vis Spectrophotometry Manual Product Overview 3 18 201...

Page 2: ...g the MenU BUTTON 6 3 Features of the Turret 6 8 A GENERAL DESCRIPTION 8 B CUVETTE HOLDER 9 C CUVETTE Z HEIGHT 9 E COOLING WATER FOR THE PELTIER 10 F DRY GAS TO MINIMIZE CONDENSATION 11 G EXTERNAL TEM...

Page 3: ...spectrophotometer 2 TC 1 Temperature Controller 3 power cable 4 USB cable 5 6 magnetic stir bars 6 vinyl tubing to connect water and gas B DEPENDING ON YOUR ORDER THE SHIPMENT MAY ALSO CONTAIN BATH 1...

Page 4: ...to the back of the TC 1 Temperature Controller on the 15 pin connectors labeled Sample and reference 5 Plug the power cable into the back of the TC 1 Temperature Controller and into a wall socket The...

Page 5: ...age push the left arrow once to get to the Set Position page Once there use the top and bottom arrows to choose a position e g position 5 Press the SET button The cuvette holder will move to position...

Page 6: ...lowly When it has locked onto a new target temperature the green LED will remain lit A rapidly flashing red LED usually means a loose electrical cable or inadequate water flow for the Peltier unit Not...

Page 7: ...own arrows and press SET The page will update to show the Current rate in min to be the same as the Ramp rate With ramping set turning on temperature control will generate a linear ramp to the target...

Page 8: ...to six cuvettes The center of the tower is hollow permitting light to pass into the entrance hole through the tower through a cuvette and exit the opposite side from the entrance A plastic cover enclo...

Page 9: ...be light tight CAUTION Accidently attaching a water line to the dry gas barb will severely damage the instrument Attach the lines with care B CUVETTE HOLDER The Turret 6 holds standard 10 x 10 mm cuve...

Page 10: ...tap for brief use You will need a flow of 100 300 ml minute This flow should require a pressure of about 3 5 psi 0 2 0 3 bar Do not exceed an input water pressure of 25 psi 1 7 bar as damage may occu...

Page 11: ...to temperature changes Also the fine coiled wires of the YSI 423 conduct very little ambient heat to the thermistor at the end Excellent Series 500 probes can be obtained with diameters less than a mm...

Page 12: ...he temperature controller is not receiving a reasonable response from the sensor on the heat exchanger Either the sensor has failed or a cable is not making a good connection E8 inadequate coolant war...

Page 13: ...10 C temperature precision 0 05 C cuvette size outside dimensions 12 5 x 12 5 mm minimum cuvette height 30 mm magnetic stirring speed 0 to 1700 rpm cuvette z height 8 5 or 15 mm specified when orderin...

Page 14: ...be established through a USB connection between the computer and the controller In this case the controller includes electronics which convert the USB connection to a serial communications port Howeve...

Page 15: ...stirrer speed F1 MS 300 The lowest stirrer speed is 300 rpm F1 SS S 1000 Set stirrer speed to 1000 rpm F1 SS S 0 Turn stirrer off and set the speed to zero F1 SS Turn stirrer on and set it to the most...

Page 16: ...us F1 IS R The controller has been powered off and back on again 7 Current Sample Temperature F1 CT What is the current temperature of the holder F1 CT 22 84 The current temperature is 22 84 C F1 CT 3...

Page 17: ...n F1 PA Start automatic reporting of probe temperature every temperature increment set by the command above F1 PT 30 5 The current probe temperature is 30 5 degrees F1 PA Stop automatic reporting of p...

Page 18: ...l send a response of the form F1 TT as a notification of the end of the ramp This response can be blocked at any time by previously sending the command F1 TT see command class 5 11 Heat Exchanger Temp...

Page 19: ...buttons and the display on the front When a computer program is in use the TC 1 will automatically send reports to the program whenever a manual change has been made The commands in this section are i...

Page 20: ...20 R1 TT R Stop reporting manual changes for the reference target temperature R1 RR R Stop reporting manual changes for the reference ramp...

Reviews: