9/17/2020
Page 18 of 44
QNW TC 1 User’s Guide
Set Stirring:
To turn on magnetic
Stirring
, use the up and down arrows to choose an approximate stirring
speed between 200 and 2000 rpm. Press
SET
to set the speed and initiate stirring. The page will update to
show the
Current
value to be the same as
Stir
speed
.
Discontinue Stirring:
Press
SET
while the
Stirring
is on (shown by nonzero
Current
rpm) to turn the
stirrer off. The page will update to show the
Current
speed to be
Off
.
Set Ramping
Figure 8
.
TC 1
Set Ramping page
Set Ramping:
To perform a temperature ramp, set the
Ramp
rate using the up and down arrows, and press
SET
. The page will update to show the
Current
rate in °/min to be the same as the
Ramp
rate. With ramping
set, turning on temperature control will generate a linear ramp to the target temperature. When the sample
holder reaches the target temperature, no further temperature change occurs, although the
Current
rate will
remain as set.
The fastest possible ramp is determined by how fast the cuvette holder is able to reach the target
temperature without ramping. Attempting to ramp too quickly, especially at high and low temperature
extremes, will result in a nonlinear ramp. The slowest ramp that may be set on this page is 0.09 °C/minute.
(If needed, much slower ramps may be set through software commands.)
Discontinue Ramping:
Press
SET
while the
Ramp
is on (shown by nonzero
Current
°/min) to turn the
Ramp
off. The page will update to show the
Current
rate to be
Off
.
3.3.
External Computer Control
All functions may be accessed either through a Serial (RS 232) or a USB located on the back of
TC 1
Temperature Controller
. You may write your own program or purchase the QNW application program
T-
App
.
T-App
will plot temperatures of the probe, cuvette tower, and even the Peltier element heat exchanger
vs
time. It will also permit you to set up temperature ramps. If you wish to do your own programming, please
see the Appendix for communication instructions and the set of text commands that may be used and
responses to the commands.
Set Ramping
Ramp = 0.55°/min