4. Decommissioning 

 

  Switch off the unit and disconnect the power plug. 

  Remove the pipes/hoses from the unit. 

  Draining and cleaning of the unit by qualified personnel.  

 
 

5. Temperature setting  

 

Thermostatic operation: adjust the desired set-point temperature at the controller (5). 
 
 

 
                    Do not operate the unit when it is open with the lid removed!  
                    Make sure that the lid is fastened before switching the unit on! 

 
 
 

For ice bank building 

  Ice bank will be build approx. 2 hours after starting the unit 

  Building  an  ice  bank  requires  filling  the  unit  with  clear  water  only,  without  Glycol  or 

remnants of Glycol in the water system  

  Recommended set-point temperature for using an ice bank: -1,0 °C (30,2 °F) 

 

ATTENTION: 

to  avoid  damage  on  the  unit  (freezing  of  items  such  as  pumps) 

make  sure  that  during  the  ice  bank  building  all  items  are  removed  from  the 
water tank

 

=> Never let the working machine unattended over a longer time

 

 
For required cold water temperatures down to -6 °C (21,2 °F) 

  Using the MiniChilly to produce cold water below the freezing point down to -6 °C 

(21,2 °F) necessitates the use of a glycol-water mixture (min. 30 % propylene glycol, 
max. 35 % propylene glycol in the water system) 

  No ice bank building is then possible due to the anti-frost qualities of the glycol 

 

Summary of Contents for MINICHILLY SB

Page 1: ...NICHILLY SB COMPATIBLE 115V WTG Quantor GmbH Europa Allee 53 D 54343 F hren Phone 49 0 6502 999 95 0 Fax 49 0 6502 999 95 75 eMail info quantor technology Homepage www quantor technology GB Attention...

Page 2: ...4 1 6 Service 4 2 Intended use 4 3 Commissioning 5 3 1 Connection 5 3 2 Chart 6 4 Decommissioning 7 5 Temperature setting 7 6 Cleaning of the cooling unit 9 7 Troubleshooting 9 8 Noise emission 10 9 T...

Page 3: ...clean from any dirt fiber etc Make sure to connect the unit to the required supply voltage Protect the unit from moisture no fluids may enter the electric parts Observe the warning and safety informa...

Page 4: ...consulted with the manufacturer The unit may only be stored in a dry dust free environment at temperature of 0 C to 40 C 1 5 Electrical installation All electrical installations are to be carried out...

Page 5: ...used to connect the plastic tubes Such kits are not provided with the unit If used please connect them according to the instructions of the manufacturer 4 Close the lid Do not operate the unit when i...

Page 6: ...omp 1 Water level gauge and tank emptying 4 Thermostatic controller 2 Cover fastener 5 Small pipes for water return flow 3 Connections 3 8 5 Expansion valve 8 Water tank 6 Condenser 9 Evaporator cooli...

Page 7: ...the unit with clear water only without Glycol or remnants of Glycol in the water system Recommended set point temperature for using an ice bank 1 0 C 30 2 F ATTENTION to avoid damage on the unit free...

Page 8: ...water mixture approx 30 propylene glycol Please pay attention that the water bath temperature must not fall below 6 C and must not exceed 30 C Controller activation for a minimum set point down to 6...

Page 9: ...Starting device at the compressor is defect Interference of the compressor interwinding fault Establish main connection plug the unit Fill in water Switch on the Thermostat Change the Thermostat Exch...

Page 10: ...s steel Mobile on wheels 10 Declaration of conformity We WTG Quantor GmbH D 54343 F hren declare in sole responsibility that the cooling units MiniCHILLY 05 MiniCHILLY 09 and MiniCHILLY 17 which this...

Page 11: ...11...

Page 12: ...E D I E N U N G S A N L E I T U N G I N S T R U C T I O N M A N U A L I N S T R U C T I O N S D E S E R V I C E M A N U A L D E I N S T R U C C I O N E S Doc ID 2018 06 SBComp115V www quantor technol...

Reviews: