7
4. Decommissioning
Switch off the unit and disconnect the power plug.
Remove the pipes/hoses from the unit.
Draining and cleaning of the unit by qualified personnel.
5. Temperature setting
Thermostatic operation: adjust the desired set-point temperature at the controller (5).
Do not operate the unit when it is open with the lid removed!
Make sure that the lid is fastened before switching the unit on!
For ice bank building
Ice bank will be build approx. 2 hours after starting the unit
Building an ice bank requires filling the unit with clear water only, without Glycol or
remnants of Glycol in the water system
Recommended set-point temperature for using an ice bank: -1,0 °C (30,2 °F)
ATTENTION:
to avoid damage on the unit (freezing of items such as pumps)
make sure that during the ice bank building all items are removed from the
water tank
=> Never let the working machine unattended over a longer time
For required cold water temperatures down to -6 °C (21,2 °F)
Using the MiniChilly to produce cold water below the freezing point down to -6 °C
(21,2 °F) necessitates the use of a glycol-water mixture (min. 30 % propylene glycol,
max. 35 % propylene glycol in the water system)
No ice bank building is then possible due to the anti-frost qualities of the glycol