
7
QNET LABVIEW HINTS
7.1 SCALING SCOPES
This section describes a handy method of changing the x or y axis in a LabVIEW scope using
QNET
_
DCMCT
_
Swing
_
Up
_
Control
VI as an example. Read the steps below to reduce the y-axis range of the
Angle (deg) scope shown in Figure 7.10 in order to see the blue trace more up close.
Figure 7.10: Scope needs to be scaled
1. As illustrated in Figure 7.11, to decrease the positive range of the scope down to 40, double-click on '100' in
the y-axis, type in '40', and press ENTER.
Figure 7.11: Scope needs to be scaled
QNET MYOELECTRIC- User Manual
14