34
This device complies with Industry Canada license-exempt RSS standard(s).
Operation is subject to the following two conditions: (1) this device may not
cause interference, and (2) this device must accept any interference, including
interference that may cause undesired operation of the device.
Le présent appareil est conforme aux CNR d’Industrie Canada applicables
aux appareils radio exempts de licence. L’exploitation est autorisée aux deux
conditions suivantes : (1) l’appareil ne doit pas produire de brouillage, et (2)
l’utilisateur de l’appareil doit accepter tout brouillage radioélectrique subi,
même si le brouillage est susceptible d’en compromettre le fonctionnement.
The following is information on potential electromagnetic interference and
advice on how to avoid or minimize such interference.
Guidance for Electromagnetic Emissions:
Appendix B. IC Statement
Appendix C. Advice for Electromagnetic Interference
QT ECG system is intended for use in the electromagnetic environment
specified below. The customer or the user of QT ECG should assure that it is
used in such an environment.
Emissions test
Compliance
Electromagnetic environment –
guidance
RF emissions
CISPR 11
Group 2
QT ECG must emit electromagnetic
energy in order to perform its intended
function. Nearby electronic equipment
may be affected.
RF emissions
CISPR 11
Class B
QT ECG is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies