Rev. 3-11
12
NOTE: To vary the
intensity remotely using
a variable DC power
supply (0-5V) instead of
a 10 K potentiometer,
connect positive to pin
8 and negative to pin 6.
9-PIN D-SUB CONNECTOR
Pin No.
Description
1
Not connected
2
Not connected
3
Not connected
4
Enables connection to a frequency counter.
5
Enables connection to an external power monitor (5 mv = 1 watt)
6
Ground
7
Energizes the ultrasonics when connected to ground.
8 and 9
Enables the intensity to be remotely adjusted using an external 10k
potentiometer.
Summary of Contents for Sonicator