background image

  

 

 

Page 9 of 32 

QSA Global Inc.

 

 40 North Avenue Burlington, MA 01803 

 888.272.2242

 

 781.272.2000

 

 F: 781.359.9179

 

 qsa-global.com 

 

Operating Conditions 

OpenVision

 

HD is designed for applications where the equipment will not be exposed to harsh handling or environmental 

extremes. See Equipment Specifications for more information. 

X-Ray Training 

It is required that OpenVision

 

HD operators are properly trained and qualified to perform radiographic inspections. This 

manual does not address radiographic inspection techniques or procedures. 

Precautions 

The  OpenVision  HD  system  is  designed  for  portability  and  is  ruggedized  for  industrial  environments.  However,  it  does 
contain fragile components. Use care to avoid strong vibrations and shock conditions (e.g. equipment drops or transportation 
outside the approved case) to the main housing, detector housing, and display(s). 

CAUTION

 

Do  not  attempt  to  open  or  modify  the  detector  housing,  main  housing,  or  X-ray  tube.  Opening  any  of  these 
components will void the system warranty and may expose the operator to a high voltage shock hazard. 

 

CAUTION 

Do not allow liquids to come into contact with any part of the OpenVision HD system. The main housing and 
detector housings have been made water resistant, but they are not “waterproof”. Appropriate care should be 
taken to protect all cabling and connectors to ensure reliable and safe operation of the OpenVision HD system. 

 

 

 

 

WARNING: Class 2 Target Laser 

Because the blink reflex (glare aversion response to bright lights) will limit the exposure to no more than 0.25 
seconds, Class 2 laser pointers are considered to be safe. Furthermore, Class 2 lasers do not harm eyes unless 
a person deliberately stares into the beam, making laser protective eyewear not necessary. A Class 2 laser is 
not a skin or materials burn hazard. NEVER aim any laser towards an aircraft or vehicle in motion. 

 

Summary of Contents for OPENVISION HD OVHD-NDT-70

Page 1: ...OPERATING MANUAL MAN 067 REVISION A OPENVISION HD OVHD NDT 70 HARDWARE MANUAL...

Page 2: ...obal Inc 40 North Avenue Burlington MA 01803 888 272 2242 781 272 2000 qsa global com Operations Manual MAN 067 March 2023 Page 2 of 32 Revision Record Revision Date Description A 03 2023 Production R...

Page 3: ...meters must be employed to avoid potentially dangerous levels of radiation exposure Take advantage of the three basic radiation protection methods to minimize radiation exposure Time Spend less time n...

Page 4: ...ting To Know Your Openvision HD 10 What s Included 10 System Overview 12 Main Housing 13 HDMI Monitor 14 Trigger Assemblies 15 Principles Of Operation 16 Radiographic Quality 16 Sources X Ray Gamma Ra...

Page 5: ...Monitor Shipping Weight 16 lbs 7 2 kg 2 lbs 0 9 kg 50 lbs 23 kg Beam Collimation 18 Horizontal 10 Vertical Startup Time 30 seconds Shutdown Time 5 seconds Battery Life Continuous Standby 2 hours 3 ho...

Page 6: ...01 2015 and meets the following standards when used as directed ANSI HPS N43 5 2005 R 2013 Ionizing Radiation Regulations 2017 ISTA 3A Over the Road vibration standard MIL STD 810 Method 514 Annex C C...

Page 7: ...Always perform the daily inspection of the system prior to use Any foreign material dirt mud ice etc must be removed before using the system Inspect clean and test the equipment as described in this m...

Page 8: ...NING ACCIDENTAL RADIATION OCCURANCES An Accidental Radiation Occurrence is defined by Part 21 CFR 1000 3 as a single accidental event or series of accidental events that has have resulted in injurious...

Page 9: ...detector housing and display s CAUTION Do not attempt to open or modify the detector housing main housing or X ray tube Opening any of these components will void the system warranty and may expose the...

Page 10: ...r Openvision HD What s Included OpenVision HD is shipped complete in a custom hardsided case Some components are packaged and placed in bags or located loose in the case Figure 1 OVHD NDT As Shipped B...

Page 11: ...Allows connection from Ethernet to OpenVision HD or Tablet 5 ELE116 USB 3 0 Adapter Connection from OpenVision HD USB A to Ethernet Adapter 6 Monitor batteries and charger 7 USB charger cable Allows...

Page 12: ...age 12 of 32 System Overview Figure 3 OpenVision HD Assembly Description Item Part No Description 1 515100 OpenVision HD Main Housing 2 515200 OpenVision HD Detector 3 515300 OpenVision HD C Arm Assem...

Page 13: ...ee M18 Battery 2 CPU Power Indicator 3 313056 Controller Key 4 415100 E01 System Status Dialight Indicator 5 RAI002 Picatinny Rail Short 6 Targeting Camera 7 Targeting Laser 8 X Ray Collimator 9 41510...

Page 14: ...itor Figure 5 and Figure 6 is a lightweight 7 inch HDMI display It can be connected to either of the two HDMI ports provided on the OpenVision HD via the HDMI cable provided The monitor comes configur...

Page 15: ...com Trigger Assemblies OpenVision HD comes standard with an adjustable trigger handle Item 515500 The trigger handle can be used for X ray firing and image video capture Refer to Figure 9 for the con...

Page 16: ...ty Curie Becquerel for radioisotopes Radiographic Quality Radiographic quality depends on an image s photographic density contrast and geometric definition distortion properties Proper energy and inte...

Page 17: ...to detector housing The C callout in Figure 10 is positioned in this approximate area Figure 10 OpenVision HD Output Boundary Distances Figure 10 shows boundary distances for 2 0 75 and 0 25 mR hr Ta...

Page 18: ...7 70 G 1 10 1 10 H 4 40 BKGD I BKGD BKGD J 2 20 BKGD K BKGD BKGD Figure 11 Example Scatter Radiation Survey Figure 11 shows an example survey of backscatter radiation user exposure for a typical samp...

Page 19: ...terial and energy level It can potentially increase the dose to the operator s extremities The best way to minimize backscatter is to use the lowest kV mA combination to achieve acceptable images Use...

Page 20: ...ngton MA 01803 888 272 2242 781 272 2000 qsa global com Operations Manual MAN 067 March 2023 Page 20 of 32 Figure 13 Example of Backscatter Figure 14 OpenVision HD with backscatter shield properly pos...

Page 21: ...a matter of seconds by following these steps Unlock main housing by pressing red circular release button and unfold to operating position Figure 15 Unlock Main Housing WARNING When the System Status I...

Page 22: ...42 781 272 2000 qsa global com Operations Manual MAN 067 March 2023 Page 22 of 32 Repeat action for detector side Figure 16 Unlock Detector Side of C Arm Position handles attach and position pistol gr...

Page 23: ...ed in multiple ways to suit different applications Figure 18 shows one such configuration The grip is moved to the top tube and the D handle is moved to the main housing Figure 18 Optional Handle Conf...

Page 24: ...lobal com Operations Manual MAN 067 March 2023 Page 24 of 32 Turn monitor ON located on upper right corner of monitor Figure 20 Monitor Power Button Install Trigger Switch by pushing up and twisting u...

Page 25: ...ON and allow time for system to initialize System status indicators located on either side of main housing will turn GREEN when system is ready Figure 22 Indicators and Power Switch Details Verify kV...

Page 26: ...nVision HD system HUD specifications Ballistic Protection Standards U S MIL SPEC MIL PRF 31013 ANSI Z87 1 U S Federal OSHA CE EN 166 FTKN Temperature Storage 4 F to 140 F 20 C 60 C Operating 32 F to 1...

Page 27: ...nditions Figure 25 Example image output for HUD 1 Micro USB charging port 2 Video input 3 Output to heads up display 4 HDMI input 5 Power 6 Reduce brightness 7 Increase brightness 8 Toggle aspect rati...

Page 28: ...Assembly Item 515500 and a non trigger Handle Item 515400 The Trigger Handle and Handle can be swapped to multiple different mounting points depending on the user s preference Figure 27 Item 515500 F...

Page 29: ...ific attention should be paid to the following M18 battery and its mount on top of Main Housing Ensure both are clean and in sound shape HDMI cable Inspect for any cracks or wear Allen thumb screws Ch...

Page 30: ...does not clear error contact your QSA Global Inc representative for repair options 003 X ray Tube Failure X ray tube is not communicating or has malfunctioned If restart does not clear error contact...

Page 31: ...ilding 12 Sunrung ro 103 Gil Kangnam Ku Seoul 06145 South Korea 82 2 558 3335 All goods and services are sold subject to the terms and conditions of QSA Global Inc A copy of these terms and conditions...

Page 32: ...ge 32 of 32 Appendix 1 Parts List Part Description OVHD NDT 70 OpenVision HD System Complete 515400 OVHD Handle Assembly 415440 OVHD Lightweight Monitor Assembly 415600 OVHD Backscatter Shield Assembl...

Reviews: