background image

To start recorder for Holter

Step 1 

  Insert formatted SD Card (using Q200 format) and battery into recorder.

 

Step 2 

  Screen will display “Q200/HE” and “QRS Diagnostic” information.* Press ENTER to continue to main 

menu. Erase memory if prompted to do so.

Step 3 

  If desired, adjust Settings by using arrows –  

and 

 – to move the cursor and the ENTER button 

to select.  Use arrows to navigate to General Settings menu and to adjust entries.  Press ENTER to 
select and return to menu.

Step 4

 

   From main menu go to New Patient screen to verify or enter Patient ID. At least one character must 

be entered for Patient ID.

Step 5

 

  Use arrows and ENTER to input ID and press EVENT when finished.  (Hold down ENTER button to 

backspace.)  Recorder will now start automatically after 10 minutes or by pressing EVENT button for 
3 seconds.  ECG signal and quality will appear on screen.

Step 6

   To remove card after recording, gently push inward to release. Never pull card out of slot, as it may 

damage the recorder.

*Note:

 

  If screen displays a 15 second countdown, you will need to interrupt by pressing ENTER,

,

, and 

EVENT buttons, in that order.

Messages

Lead Loose

 – Occurs when the patient is not hooked 

up or if there is a problem with the hook up.  The 
problem may be with an electrode, a lead, or the cable 
that connects the leads to the recorder. The message 
will continue to flash for about 10 seconds after 
corrected. When corrected, the ECG, then Time-of-day 
will appear on the screen. If you choose, you can turn 
this error message off via the main menu.

Battery LOW

 or 

 

FAILURE

 

– Put in new battery before 

starting recorder. 

Erase memory YES/NO

 

– If the SD card has been 

used for a previous patient, you will need to erase now.   
If the card should not be erased as it has ECG data that 
you do not want to lose, you will need to remove  
it and put in a new formatted SD card.

 

SD Card errors

 

– SD Cards must be formatted using 

your LX Holter software. Refer to your Operator’s 
Manual for details on SD Card errors and how to format 
and correct. 

General Settings

Contrast

 

Adjusts LCD screen contrast

Lead Loose

  

Enables/Disables “Lead Loose”  
error message

Event Marker

 

Denotes event location on ECG 

Key Mode

  

“Quiet” turns off key beeping.  
“Delayed” prevents accidental events 
by requiring patient to hold button 
down for several seconds to record an 
event

Rec Type

 

 Switch between Holter and Event only 
recording modes.

Menu Lock

  

Locks settings on the recorder 
To lock recorder enter “217” 
To unlock the recorder enter “151”

Language

 

Choose from multiple languages

Hi Res

 

 Enables/Disables high resolution 
recording (Holter only.)

 

QRS Diagnostic 

 

6901 E. Fish Lake Road, Suite 188 
Maple Grove, MN, USA 55369
800.465.8408    |    +1.763.559.8492    |    Fax +1.763.559.2961
[email protected]    |    www.QRSdiagnostic.com

0086

Release Date: 04/12

Advena Ltd

33 Bridge St.

Hereford HR4 9DQ

UK

© QRS 

600005-00A   

Q200/HE Quick-Start Guide for Holter

Reviews: